DMM OPERATOR’S MANUAL

A960T

1. Overview

The DMM are characterized at slim size, portable, stable

performance and anti-dropping capacity. Using 3½ digits LCD

monitor with character 15mm high, they offer clear readings.

With overall circuitry design centering on large-scale IC A/D

converters in conjunction and over-load protection circuit, the

meters give excellent performance and exquisite making as a

handy utility instrument.

The meters can be used to measure DC & AC voltage, DC

current, resistance, positive diode voltage fall, hFE parameters

for transistor and Continuity.

2. Panel Layout

1

LCD

Display:

digits,

character 15mm high

2

Back Light Button Switch:

Press this button to switch on

back

light

If

the

dark

circumstance light makes the

reading

difficulty

when

measuring, the light will be

automatically turned off in 5

seconds. Press again to switch

it on again. If the battery is in

weak power, the light will be

dimmed.

3

Data-hold Switch (HOLD)

4

Power

5

Rotary Switch: use this switch to select functions and

ranges

6

“10A”Input Jack

7

hFE test jack

8

“COM”Input Jack

9

“VΩmA” Input Jack

3. Safety Information

3-1 The meters are designed according to IEC-1010 concerning

electronic measuring instruments with an over-voltage

category (CAT II) and pollution 2.

3-2 Follow all safety and operating instructions to ensure that

the meter is used safely and is kept in good operating

condition.

3-3 safety symbols:

important safety information, refer to the operating manual.
Dangerous voltage may be presence.

Double insulation (protection Class II)

4. Special Cautions for Operation

4-1 The meters can be safe only according to standard

procedures when used in conjunctions with the supplied

test leads. To replace damaged test leads with only the

same model or same electric specifications.

4-2

To aveid risk of electric shock, do not use the meters

before the cover is in place.

4-3 The range switch should be at right position for the testing.

4-4 To avoid electric shock and damaging the instruments, the

input signals are forbidden to exceed the specified limits.

4-5

When measuring TV set or switched power, attention

should be paid to the possible pulses that may bring

destruction to the circuit.

4-6

Range switch position is forbidden to be changed at

random during measurement.

4-7 Take caution against shock in the course ot measuring

voltage higher than DC 60V & AC 30V.

4-8 Protection fuse should be replaced only with same type

and same specification.

5. GENERAL SPECIFICATIONS

5-1 Max Voltage between input terminal and Earth Ground: CAT

II 600V

5-2 Over-range Indication: display “1” or “-1” for the significant

digit.

5-3 Automatic display of negative polarity “ - ” .

5-4 Low Battery Indication: ‘ ’ displayed

5-5 Max LCD display:

1999 (3

1

/

2

digits)

5-6 Fuse protection:

F-200mA/250V (Ø5x20mm)

5-7 Power Supply:

9V battery, 6F22 or NEDA 1604

5-8 Operating Temp.:

0

to 40

(relative humidity <85%)

5-9 Storage Temp.:

-10

to 50

((relative humidity

<85%)

5-10 Guaranteed precision Temp.: 23±5

(relative humidity <85%)

5-11 Dimension:

69x138x31mm (with holster)

5-12 Weight: approx: 170g (including battery)

6. Testing Specifications

Accuracy is specified for a period of year after calibration

and at 18

to 28

(64

to 82

) with relative humidity to

75%.

6-1 DC Voltage

Range

Resolution

Accuracy

200mV

0.1mV

±(0.5% of rdg + 2 digits)

2V

1mV

20V

10mV

200V

100mV

±(0.8% of rdg + 3 digits)

500V

1V

-- Input Impendence: 1MΩ

-- Overload protection: 250V for 200mV range, effective DC or

AC 600V for other ranges

6-2 AC Voltage

Range

Resolution

Accuracy

200V

0.1V

±(1.2% of rdg + 10 digits)

500V

1V

-- Frequency Range: 40 to 400Hz

-- Response: average, calibrated in rms of sine wave

6-3 DC Current

Range

Resolution

Accuracy

200μA

0.1μA

±(1.0% of rdg + 2 digits)

2mA

1μA

20mA

10μA

200mA

100μA

±(1.5% of rdg + 2 digits)

10A

10mA

±(3.0% of rdg + 5 digits)

-- Overload protection: F 200mA/250V fuse

Note:

10A range: not fused

6-4 Resistance

Range

Resolution

Accuracy

200Ω

0.1Ω

±(0.8% of rdg + 3 digits)

2KΩ

20KΩ

10Ω

200KΩ

100Ω

2MΩ

1KΩ

±(1.0% of rdg + 3 digits)

-- Over-load protection: 250V effective value

6-5 Transistor hFE Test

Range

Test Range

Test Current / voltage

NPN & PNP

0-1000

I

b

=10µA / Vce=3V

6-6 Diode Test

Range

Resolution

Function

1mV

Display: read approximate

forward voltage of diode

-- Over-load Protection: 250V effective value

-- Forward DC current: approximate 1mA

-- Reversed DC voltage: approximate 3.0V

6-7 Continuity

Range

Function

Built-in buzzer will sound if resistance is

less than 60Ω±30Ω

-- over-load protection: 250V effective value

--open circuit voltage: approximate 3.0V

6

-8 Temperature

Range

Resolution

Accuracy

-20

-750

1

±(1.0% of rdg + 2 digits)

-- over-load protection: 250V effective value

7. OPERATING INSTRUCTIONS
7-1 Attention before operation

7-1-1 Check 9V battery. If the battery voltage is less than 7V,

display will show “

”, the battery should be replaced at

this time to ensure measuring precision.

7-1-2 Pay attention to the “

” besides the input jack which

1

2

3

4

5

6
7

8