INSTRUCTION MANUAL

FOR

ULTRASONIC NEBULIZER

MODEL CUN60

Eng 1

 -

Summary of Contents for CUN60

Page 1: ...English Espa ol Portugu s Deutch Fran ais INSTRUCTION MANUAL FOR ULTRASONIC NEBULIZER MODEL CUN60 Eng 1...

Page 2: ...d Silent operation thanks to the ultrasonic vibration technology 3 speed nebulization rate control provides optimal treat ment for all kinds of medication with different density Built in timer automat...

Page 3: ...rilize all necessary parts as described on page 8 Do not handle the power cord with wet hands or im merse the AC adapter An electrical shock may result The medication cup mouth nozzle and air filter s...

Page 4: ...n or manufacturer of medica tion CLEANING Never immerse the equipment in water It may dam age the unit Clean all necessary parts after each use as described on page 8 STORAGE Do not store the unit und...

Page 5: ...PARTS DESCRIPTION Mouth nozzle Filter cover Air filter Top cover Medication cup Fan Water reservoir Low battery indication lamp Main body Nebulization rate Indication lamps ON OFF button Nebulization...

Page 6: ...over 30 C Caution 1 Do not use solutions other than water for the res ervoir It may damage the transducer 2 Make sure the water level is between 2 lines in the reservoir OPERATION 2 FILL MEDICATION CU...

Page 7: ...at high medium or low speed which will be illuminated by corresponding indication lamp b Begin treatment according to your physician s instruc tions Caution Stop operation immediately if any abnormali...

Page 8: ...e top cover b Pour remaining medication out of cup c Drain water from water reservoir Caution Do not touch water it is hot after use 2 CLEANING a Nose piece mouth nozzle extension tube mask top cover...

Page 9: ...le size 5 m MMAD Medication cup capacity Approx 8 ml max Transport storage condition 10 50 C 45 85 Safety classification Type B apparatus Class II insulation Auto shut off timer Approx 10 min Dimensio...

Page 10: ...filled to the correct level If using with battery only check that battery pack is firmly attached to the unit Check that battery is fully charged Check that there s medication in the cup Check that t...

Page 11: ...battery is approx 2 years However battery life differs according to the usage and charg ing conditions To save battery power and avoid accidental operation always turn the switch at bottom at OFF CHAR...

Page 12: ...operation is finished turn the switch at bottom to OFF CHARGE position to save battery power d To remove battery pack press the grooved area and release it from nebulizer by sliding down Note The swit...

Page 13: ...e Minimum 24 hours first time 4 12 hours thereafter depending on duration of usage Capacity 480 mAH 12VDC approx 30 minute continuous use Operating temperature 0 45 C 32 112 F Storage temperature 18 4...

Page 14: ...EN is a registered trademark of Citizen Watch Co Japan Consumer Electronics Division CBM Bldg 5F 5 68 10 Nakano Nakano ku Tokyo 164 0001 Japan Phone 03 5345 7430 Fax 03 5345 7431 E mail info con jcbm...