Suggested Electrode 

Placement Chart

The sticky electrode pads need

to be placed in the correct

position for maximum pain relief.

It is preferable to obtain

assistance from either your

birthing partner or medical

adviser when positioning the

pads. Please read the following

instructions carefully:

1.

Ensure unit remains switched

off. Carefully peel the pads from

the plastic backing using your

forefinger and thumb. Do not pull

on the lead wires. 

2. 

Position the top two pads

(connected to the left socket

CH1) approximately 4cm either

side of the spine and 4-5cm

above the imaginary navel line as

seen in the diagram. 

3.

Position the bottom two pads

(connected to the right socket

CH2) approximately 4cm either

side of the spine and 4-5cm

below the imaginary navel line as

seen in the diagram.

Reference: Cluett E. (1994) SRN

RM ADM PGCEA. Analgesia in

Labour: A review of the TENS

method. Professional Care of

Mother and Child. Mar;4 (2): 

50-2.

Bra line

Navel line

Panty line

Using Your Obi TENS +

a)

You are now ready to start

using your Obi TENS +. Switch

the unit on by pressing the

on/off button. The Obi TENS+ will

always start up in BURST 1 mode.

b)

Begin by pressing the

CH1+ key. Each time you press

this key, a small ‘half-moon’

shape will appear on the screen

increasing in number and size as

the intensity is increased. You

will begin to feel a pulsating

sensation which will burst in time

with the small arrow flashing on

the screen. 

Ensure that you

increase the intensity slowly.

c)

Repeat the previous step with

the         CH2+ key. You will now

feel a sensation in both sets of

pads.     

d) BURST 1 (use between

contractions):

This is the first

mode that you will use. It feels

like a bursting, tingling sensation

i.e. the pulses are off and on

periodically. This mode will help

to promote those all-important

natural pain-killing chemicals

known as ‘endorphins’ and

‘encephalins’.

e) BOOST 1 (use during

contractions):

When

experiencing a contraction,

simply press the ‘boost’ button

located on the left side below the

screen. The Obi TENS+ will then

switch automatically into the

BOOST 1 mode for that extra

surge of power required to

combat the pain during

contractions. This will feel like a

continuous sensation.

f )

Once the contraction has

passed, simply press the

‘BOOST’ button. The Obi TENS+

will return to the original BURST

1 mode. Continue this cycle

during your early stages of

labour.

Using the Between Contraction

Timer (BCT) timer 

When in BURST mode, the unit

automatically measures your BCT.

This appears at the bottom of the

display. When you switch to

BOOST, the last BCT will remain at

the bottom of the display. The

BCT will always refresh when re-

entering the BURST mode from

the BOOST mode and will record

the next BCT.

N.B. The Obi TENS+ starts up in

BURST Mode. The BCT only

appears after one cycle of

pressing the Boost Button on

and off.

Recalling contraction history

To recall a list of your last 5

‘BCTs’ press the CH1-       and

CH2-        buttons

simultaneously. The 1st symbol

you will see is a single moon

which is the last BCT. 

Use CH2+         to reveal up to 5

BCTs            . Use CH2-         to

review more recent BCTs.

Recalling the contraction history

will not disable functionality of

the unit. To return to the main

screen press the CH1-       and

CH2-       buttons simultaneously.

N.B. To avoid losing contraction

history, do NOT switch unit off.

Remove the battereis if you wish

to reset the contraction history.

The Mode Button

There are two modes available

on the Obi TENS+, each with a

BURST and BOOST function. 

Mode 1 (Burst 1 and Boost 1)

,

which has been described

above, is primarily used for the

earlier stages of labour. 

Mode 2 (Burst 2 and Boost 2)

can be used when your labour

advances and your contractions

become more frequent. 

Press the Mode          button to

switch between modes. You will

see from the User Display, that

the number next to the BURST or

BOOST will change from 1 to 2

and vice-versa.

NB. Both modes are

interchangeable and can be

used according to your

preference, not necessarily for

the stage of labour you are in.

 Mode 2 can be operated in the

same way as mode 1.

Useful Information

1. Continue stimulation for as

long as necessary, unless you

experience discomfort.

2. If you stimulate for a number

of hours ensure the pads do

not become dry. 

If this occurs, switch off the

unit, disconnect the

electrodes, and apply a small

drop of water to them.

3. The pads can be used several

times. Replace the pads

when they stop sticking to

the skin. Replacement pads

are available from

www.babycaretens.com.

4. Always ensure that the unit is

switched off before you

remove all the leads and

pads. Do not pull the wires.

ALWAYS remove the jack plug

from its socket by holding

the plug between the

forefinger and thumb and

pulling gently.

5. The Obi TENS + can also be

used as a standard TENS unit

for relief of numerous

conditions, including after

pains and post operative

pains. It is worth bearing this

in mind if you have

a caesarean section. 

Please seek medical

advice first.

Post-Natal Pad Placement

Diagrams

KEY:

=first pair of electrodes

=second pair of electrodes.

N.B. Electrodes supplied with this unit

are all the same colour - the black and

white squares show suggested

electrode placement only.

Frequently Asked Questions

Q

When shall I start using my

Obi TENS +?

A

Use as early as you can after

the onset of labour to allow

time for your body’s pain-

killing chemicals to rise.

Q

Can I try the Obi TENS +

before going into labour?

A

Yes, place 2 pads attached to

one leadwire on your forearm

to feel the different pulse

sensations following the

instructions on page 6.

Q

Can I combine the Obi TENS

+ with other medication?

A

Yes, you can still use

analgesics such as gas and

air (entinox) or pethidine.

Q

Can the Obi TENS + be used

in hospital?

A

Yes. However, a midwife may

ask you to switch it off

temporarily.

Q

Can I use the Obi TENS + 

if I am considering a water

birth?

A

Yes, but not in the water

itself.

Troubleshooting Questions

 

Q

Why does the stimulation

not appear to be as strong

after you have used the Obi

TENS + for a while?

A

Increase the intensity, you

may have become used to a

lower setting.

Q

Why does the stimulation

feel strong but ineffective?

A

You may need to switch off

the unit and reposition the

electrode pads.

Q

Why does the stimulation

not feel sufficiently strong

even on a high setting?

A

The batteries may need

replacing.

Q

What do I do if there is no

stimulation, intermittent

stimulation, or weak

stimulation felt even with a

new battery?

A

Check electrode contact.

Check leads. 

Check batteries. 

Call our customer careline 

on +44 (0)208 5329 5959.

0120

July 2017 (1) 

 

 

Babycare TENS

108 George Lane, South Woodford,

London E18 1AD 

Tel (+44) (0)20 8532 9595

Fax (+44) (0)20 8532 9551 

www.babycaretens.com

e-mail [email protected]

Fuji Dynamics Ltd, Unit C,3/F, Manley Tower,

No828 Cheung Sha Wan Rd, HK

This manual is recyclable

Babycare TENS is part of the 

Body Clock group of companies

Guarantee

Your Obi TENS + machine is guaranteed for a period of 1 year against

manufacturer’s defects. Please record purchase details in the space

provided below for your record purposes. The guarantee does not

include leads, electrodes or battery. Applies to purchases only.

Date of purchase:

Invoice number:

2

2

1

1

Display

TN type LCD 

Operation Mode

Burst 1

Burst 2

Boost 1

Boost 2

4 treatment modes: 

Burst 1, Burst 2, Boost 1, Boost 2

Pulse (Hz)

Frequency

Pulse Width

(µs)

Description

16

150

1 Burst/sec and 

8 pulses per burst 

32

150

2 Burst/sec and 

8 pulses per burst 

80

150

Constant

100

150

Constant

- Output Voltage: 0 – 55V zero to peak at 500Ω load

- Output Waveform: Symmetrical biphasic rectangular

Automatic Power Off

5 Minutes

Output Channel

Two Channel

Output Voltage

0 – 55V (zero-to-peak) across a 500Ω load; 

adjustable in 14 steps

Output Intensity

0 – 110 mA (max.) into a 500Ω load

Output Waveform

Symmetrical bi-phasic rectangle 

Output Jacks

CE Touch-proof jacks

Load Detector

Contact detection between the device and human skin

(This feature is activated from Level 2 and up)

Battery

2 X 1.5V AAA Battery

Battery Low Detect

2.2V ±0.2V

Tolerances

All output parameters are subjected to a ±20% tolerance

unless other specified.  

Weight

52g 

Size

40 x 125 x 20mm

Period Pain

Lower Back Pain

Shoulder Pain and Frozen Shoulder

Outer and inner aspect of shoulder

Obi TENS + Technical Data