Summary of Contents for COVIDAIR S/T EASY

Page 1: ... 1 Quick Start COVIDAIR S T Easy English 1 1 docx BCD 30 04 2020 COVIDAIR S T EASY QuickStartGuide WARNING The COVIDAIR device must always be used under the supervision of a doctor or a member of the medical profession trained in ventilation ...

Page 2: ...hing ON the device 5 2 2 Switching OFF the device 5 2 3 Start and Stop ventilation 6 2 4 Locking and unlocking the touch screen 7 2 5 Navigation in menus and settings 8 2 5 1 Cockpit Page 9 2 5 2 Graphic Page 10 2 5 3 Ventilation Parameter Page 11 2 5 4 Alarms Page 13 2 5 5 Advanced Parameters Page 15 2 5 6 History Page 16 2 5 7 Display Brightness 17 2 5 8 WiFi Parameters 18 ...

Page 3: ...COVIDAIR S T Easy Quick Start Guide Page 3 Quick Start COVIDAIR S T Easy English 1 1 docx BCD 30 04 2020 ...

Page 4: ...hedevice ON The COVIDAIR device starts up as soon as the external power supply is connected The push button flashes slowly 1 Hz to signify that the device has started If the device has been switched off displaying the graphical interface and the mains supply has not been disconnected press the push button to start it again The device is ready to use when the touch screen displays the main screen T...

Page 5: ...R S T Easy Quick Start Guide Page 5 Quick Start COVIDAIR S T Easy English 1 1 docx BCD 30 04 2020 WARNING Always switch off the COVIDAIR device via the interface before disconnecting the external power supply ...

Page 6: ...n of the stop and the message Ventilation stopped is displayed Start via the main push button Press the main push button to start ventilation The ventilation starts with the parameters saved in the Parameters and Alarms tabs The message Ventilation activated is displayed and the main button flashes each time the patient breathes Press the main push button again to stop the ventilation a message th...

Page 7: ...unwanted manipulation This function is only available from the main screen To lock the touch screen press the padlock button on the main screen Once the screen is locked the following message is displayed To unlock the screen press the main push button NOTE Alarms have priority over screen locking If one of them sounds while the screen is locked it will unlock and show the alarm in a visible way ...

Page 8: ...menus and settings Navigation in the page and settings is carried out using the tabs at the top of the display The tabs are visible in each menu Main page Cockpit Graphic Display page Ventilation parameter page Alarms page Advance parameter page History page Maintenance Trained user only Brightness setting Network Settings ...

Page 9: ... cmH2O 8 IPAP Inspiratory Positive Airway Pressure in cmH2O 9 HR Heart rate indicator in beats per minute only if the pulse oximeter is used 10 SpO2 Oxygen saturation indicator in only if the pulse oximeter is used Validity from 90 to 100 11 VTt Expiratory tidal volume with leakage correction 12 MV Current Patient Volume in liters per minute 13 Graphic area with multiple displays The selection of ...

Page 10: ...COVIDAIR S T Easy Quick Start Guide Page 10 Quick Start COVIDAIR S T Easy English 1 1 docx BCD 30 04 2020 14 Display area for various devices messages 15 Screen lock button 16 Device shutdown button ...

Page 11: ...in liters per minute 2 Pressure Paw in cmH2O 3 Time scale It is also possible to zoom directly on the graphics with two fingers Displays the 60 last seconds Displays the last 30 seconds Displays the last 10 seconds 4 RR Respiratory Rate per minute 5 VTe Expiratory tidal volume with leakage correction 6 I E Inspiratory Expiratory time ratio ...

Page 12: ...tment of the EPAP expiratory pressure in cmH2O The value of the expiratory pressure is between 2 and 40 cmH2O it 1 cmH2O increments but it cannot be greater than the IPAP value If IPAP and EPAP are equal the device functions as CPAP 3 Adjustment of the IPAP inspiratory pressure in cmH2O 4 Tidal volume adjustment WARNING This value did not take care of leaks It is not recommended to use this value ...

Page 13: ...ry trigger in The higher the the more sensitive the trigger If the trigger is too sensitive too high there is a risk of self trigger 9 Inspiratory trigger sensitivity activation slider see its meaning and use at the end of the chapter 10 Setting the maximum inspiration time in seconds The value of the inspiration time is between 0 1 and 3 by 0 1s increments ...

Page 14: ...er is not used 14 Humidity activation slider see its meaning and use at the end of the chapter For each setting instruction the button allows you to decrease the value while the button allows you to increase the value by the indicated steps It is also possible to let your finger pressed in order to reach the settings faster The sliders in front of certain parameters allow you to activate or not th...

Page 15: ...he lower limit for the alarm Total expired volume VTe in milliliters The value of the lower limit for Total expired volume is between 40 and 1300 by 5ml increments 3 Setting the upper limit for the alarm Total expired volume VTe in milliliters 4 Activation slider for the Volume per minute MV alarm see its meaning and use at the end of the chapter 5 Setting the low limit for the Volume per minute M...

Page 16: ...60 by 5s increments 15 Activation slider for the Allowable leak time Leak alarm see its meaning and use at the end of the chapter 16 Setting the time limit for the Allowable leak time Leak alarm in seconds The count begins at the start of the leak The limit value for the Allowable leakage time is between 5 and 60 by 5 s increments 17 Setting the low limit for the Heart rate Fc in beats per minute ...

Page 17: ...cursor is ON the parameter is active When the cursor isOFF the parameter is not active NOTE Alarms take precedence over the screen lock or any other displayed menus If one alarm activates it is displayed in priority See also the Alarms chapter earlier in this document for details on viewing and managing current alarms ...

Page 18: ...ed out under the supervision of an authorized technician 1 Option to automatically start ventilation at power on 2 Name of the device 3 Manufacturer 4 Firmware revision 5 Graphic user interface revision 6 Device MAC address 7 Total operating time 8 Total turbine time 9 Last ventilation operating time or time since the last ventilation operating time 10 Air humidity and temperature indicator suppli...

Page 19: ...CD 30 04 2020 COVIDAIR S T Easy Quick Start Guide Page 18 18 2 5 6 HistoryPage The History menu displays the history of events that occurred during each ventilation cycle 1 History of all events since the device was switched on 2 Clear history button ...

Page 20: ...0 COVIDAIR S T Easy Quick Start Guide Page 19 19 2 5 7 Display Brightness The display brightness is adjustable at all times via the button 1 The menu opens when the button is pressed 2 Slide from left to right in order to adjust the brightness for visual comfort ...

Page 21: ... is necessary to connect a USB keyboard mouse to the dedicated connector on the back of the device 1 The menu opens when the button is pressed 2 SSID Wireless network name 3 Wireless network Password 4 Location Location of the COVIDAIR device useful during operation with supervision Example HUG GENEVA room 1822 P1 5 Server address Dedicated server for remote supervision IP or domain name 6 Status ...