Medical Compressor Nebulizer 

Version 1.0

Thank you for choosing AVICHE Medical Compressor Nebulizer. Please read this User 
Manual carefully before use and maintenance and keep it properly for future reference.

Model: AH-C10

User Manual

Summary of Contents for AH-C10

Page 1: ...or Nebulizer Version 1 0 Thank you for choosing AVICHE Medical Compressor Nebulizer Please read this User Manual carefully before use and maintenance and keep it properly for future reference Model AH...

Page 2: ......

Page 3: ...nit Operating Instructions Preparing the Nebulizer for Use Attaching the Air Tube Using the Device Care and Maintenance Cleaning after Each Use Daily Disinfecting Changing the Air Filter Caring for th...

Page 4: ...he following components Compressor Main Unit Nebulizer Kit Air Tube PVC 150 cm Air Filters Package of 2 User Manual Adult or Child Mask PVC For specific information about your own nebulizer CONSULT YO...

Page 5: ...device where the device may be exposed to flammable gas or vapors Do not use tap or mineral water in the nebulizer for inhaling purposes Clean and disinfect the nebulizer kit mouthpiece and optional...

Page 6: ...ore use Do not use in anaesthetic or ventilator breathing circuits Not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide Do not operate where oxygen...

Page 7: ...operates normally Do not use the device if the air tube is bent Do not block the air filter cover Do not alter the baffle the nozzle in the medication cup or any part of the nebulizer kit Do not add...

Page 8: ...zation be sure to wipe the face after removing the optional masks When using this device there will be some noise and vibration caused by the pump in the compressor There will also be some noise cause...

Page 9: ...n an operational failure Do not overload power outlets Plug the power cord into the appropriate voltage outlet Do not use extension cords Plug the power cord directly into the electrical outlet Unplug...

Page 10: ...wrap the power cord around the compressor main unit and power cord Components Know Your Unit User Manual including Warranty Card 1 Air Filters Package of 2 Product Inspection Certificate 1 Medical Com...

Page 11: ...izer Kit Inhalation Top Inhalation Cap Knob Inhalation Cap Medication Cup Mask Air Tube Connector Baffle Nozzle Ventilation Slots Nebulizer Kit Holder Power Switch Air Tube Connector Air Filter Cover...

Page 12: ...sed on average for more than 60 days replace it with a new one For directions on cleaning and disinfecting refer to page 16 18 For directions on changing the air filter refer to page 18 1 Plug the pow...

Page 13: ...ion on the medication cup as illustrated A Turn the inhalation top clockwise until securely closed B For type dose and regime of medication follow the instructions of your physician or licensed health...

Page 14: ...Medication may spill out of the medication cup 5 Attach the desired inhalation accessory 1 Twist the air tube plug slightly and push it firmly into the air tube connector on the upper side of the comp...

Page 15: ...id locations such as a bathroom Use the device within the operating temperature and humidity Provide close supervision when this device is used by infants children or compromised individuals Do not us...

Page 16: ...filter cover Limit use to 30 minutes at a time and allow a 30 minutes interval before using the device again When using the device the main unit may become hot Do not touch the main unit for other th...

Page 17: ...as a result of bacteria WARNING If any condensation or moisture remains in the air tube remove the moisture from the air tube Follow the directions below 1 Make sure the air tube is still connected to...

Page 18: ...ntenance Following cleaning instructions after each use will prevent any remaining medication in the bottle from drying resulting in the device not nebulizing effectively and will help prevent infecti...

Page 19: ...nstructions provided by the disinfectant manufacturer 1 Submerge the parts in the cleansing solution for the specified period 2 Remove the parts and discard the solution 3 Rinse the parts with clean h...

Page 20: ...rial for the air filter Wash the air filter cover regularly to prevent any blockage in the cover Do not boil Make sure the cover is dry before inserting the new air filter CAUTION 1 Pull the air filte...

Page 21: ...ly using a soft clean cloth Do not immerse the compressor main unit in water or other liquid Unplug the power cord from the electrical outlet before cleaning the device Do not insert any object into t...

Page 22: ...before using the device again The nozzle is blocked Clean and disinfect the nebulizer kit to remove the blockage The nebulizer kit is tilted at an incorrect angle Hold the nebulizer kit correctly Do...

Page 23: ...hase Include a letter with your name address phone number and description of the specific problem Pack the product carefully to prevent damage in transit Because of possible loss in transit we recomme...

Page 24: ...be determined by turning the product on and off the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase t...

Page 25: ...Type Power Supply Nebulization Rate Particle Size Medication Cup Capacity Appropriate Medication Quantities Operating Condition Operating Temperature Humidity Dimensions Weight Storage Temperature Hum...

Page 26: ...is standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices Medical devices manufactured by AVICHE conform to...

Page 27: ...luctuations Flicker Emissions IEC 61000 3 3 Compliance Group 1 Class B Class A Complies Electromagnetic Environment Guidance The AH C10 uses RF energy only for its internal function Therefore its RF e...

Page 28: ...70 UT 30 dip in UT for 25 cycles 3A m Note UT is the A C mains voltage prior to application of the test level 1 The test of input output lines is not applicable since they are shorter than 3 0m 3A m...

Page 29: ...stance calculated from the equation appropriate to the frequency of the transmitter Recommend separation distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum ou...

Page 30: ...t and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the mea...

Page 31: ...to 2 5GHz d 2 3 P 0 12 0 12 0 23 0 37 0 37 0 74 1 17 1 17 2 33 3 69 3 69 7 38 0 01 0 1 1 10 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended sepa...

Page 32: ...30...

Page 33: ...warranty period will be identified by extension of 13 months from the date of production 3 The followings are not covered by the warranty 3 1 The Vulnerable Consumables 3 2 The liquid leaked by the u...

Page 34: ...and free mainte nance will be provided except for force majeure and human factors 3 2 The affiliated gifts will not be covered within the warranty 3 3 Proof Purchase records shopping certificate and i...

Page 35: ......

Page 36: ...Co Ltd Address Floor 26 Building A A2 1 Jinan Pharm Valley Gangxingsan Road High tech District Jinan City Shandong Province China Service Hotline 86 400 800 9850 Website www aviche cn Email info avic...