Temperature Monitor/Logger DS-108

 

 
English User Instructions 
 
General Description

 

 

The 

DS-108

 is an 8 channel temperature monitor and logger which connects to a PC via a 

USB port, ideal for use in laboratories, workshops, classrooms and on site. It has 8 input 
channels  for  use  with  digital  temperature  probes  which  operate  between  -55C  (-67F)  to 
+125C (+257F), as well as an internal circuit/ambient temperature (a ninth channel). The 
DS-108 also has 8 digital (0/1) input channels which allow the monitoring and logging of 
the  status of switches, interlocks, pumps, compressors, fans, doors, etcetera. The digital 
input  channels  can  accept  DC  voltages  up  to  30Vdc,  voltages  from  0V  to  1.8V  are 
considered  logic  0  (off)  and  voltages  from  2.2V  to  30Vdc  are  considered  logic  1  (on).  A 
range of signal conditioning interface circuits are available to convert parameters such as 
mains voltage, light intensity, magnetic field strength to a voltage level appropriate  to the 
digital input channels (0-5Vdc). 
 

1. DS-108 hardware

 

The 

DS-108 

Is  housed  in  a  strong  ABS  case  with  8  temperature  inputs  on  the  front,  via 

3.5mm jacks and 8 digital inputs (0Vdc to 30Vdc) on the rear, via screw terminals. The DS-
108  also  has  a 5V/400mA output to power any peripheral signal conditioning circuitry. To 
communicate with the PC the DS-108 must be connected to a Personal Computer (PC) via 
a USB A to B cable.  
 

2. TempLog Software (download free from www.acksen.com)

 

TempLog runs under Microsoft

®

 Windows XP, Vista, 7, 8, 10 or 2003/2008/2012 Server. It 

will allow the user to:- 
Monitor and record data from the DS-108 on the PC, in both tabular and graphical format. 
Store the data to disk and / or export to spreadsheet format. 
Print data and/or reports. 
 

3. USB A to B cable

 

Used to connect the DS-108 to a PC. 
 

4. Carry case 

 

The carry case allows you to keep the DS-108, temperature probes, USB lead and any 
conditioning circuitry together. 
 

5. Digital Temperature Probes and Leads 

 

The standard DS-

108 ‘SET’ comes complete with 4 (four) digital temperature probes with 

3.5mm plugs, with 3m (10ft) leads, 

‘DS-108-

4

-

3

- SET

’. 

 

 
Hardware requirements 

To  run  TempLog  you  must  have  a  PC  with  Microsoft  Windows  XP,  Vista,  7,  8,  10  or 
2003/2008/2012 Server. 
 
Our  minimum  system  specification  is  a  2Ghz  Intel  Core2Duo,  2GB  RAM,  1GB  free  disk 
space, 1366x768 resolution, 24-bit colour and Windows XP. 
 
Our  recommended  system  specification  is  a  3.1Ghz  Intel  i5-3450,  8GB  RAM,  1GB  free 
disk space, 1920x1200 resolution, 24-bit colour and Windows 7, 8 or 10. 
 

Installing TempLog (download free from www.acksen.com)

 

Download  and  run  the  TempLog  installation  software  from  our  website.    Follow  the  on-
screen instructions to complete installation.  After installation run TempLog from the Start 
menu in Windows. 
 

 
 
 
 
 
 
 
 
 

 
 
 
Getting started 

The DS-108 is powered from the USB port on the PC, it normally takes around 100mA. The 
5V  output  on  the  DS-108  is  fused  to  400mA,  so  the  DS-108  should  never  take  more  than 
500mA. When you first connect the DS-108 to the PC you may have to wait until Windows 
installs the correct drivers before you can communicate with the DS-108. 
 
When  the  DS-108  is  connected  to  the  PC  it  is  automatically  powered,  signified  by  the 
flashing blue LED on the front.  
 
Open the TempLog software and if you have not run the software before, register and then 
use the ‘Detect Logger’ option to connect to the DS-108. 
 
To record the data from the DS-

108, simply press the ‘Start Recording’ button. 

 

This  document  is  produced  in  conjunction  with  the  Help  file  contained  in  TempLog,  which 
contains  a  detailed  explanation  of  all  features  and  contains  important  information  which 
should be studied prior to using this product.

 

 

Within the software you can set the units to Celsius, Fahrenheit or Kelvin. You can set 
temperature alarm levels, label and colour the channels and set many other parameters. 
You can also input details of the data recording situation, the location, the item under test, 
the experiment details. Data can be viewed in a tabular form, a dashboard or in a graph. 
 
For tests longer than 24 hrs the data is automatically saved to a new file each day at 12:00 
midnight, so tests that are longer than a day will be split into files of a maximum duration of 
24 hours; a week long test will have 7 files each representing 24 hours of data. 
 

 
Features & Benefits  

Feature 

Benefit 

Unit is small and lightweight. 

Easy to install. 

Easy to use Windows software. 

Can be used by non-technical staff. 

Accurate temperature measurement. 

Digital probes require no calibration. 

Digital logic inputs 

Allows correlation of other parameters 

 

 
The standard 

‘DS-108-

4

-

3

- SET

’ is supplied with 

(four) digital temperature probes, with 

3

leads. These probes require no calibration. The probes have a range of -55C to +125C 
which is  -67F to +257F. Simply plug the 3.5mm plugs into the relevant channel on the 
logger, start the TempLog software and your logger will transmit the readings to the PC. For 
help using the software, please read the Help File shipped with the software. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
The digital input channels are used to monitor, record and correlate the status of other external 
ancillary or related equipment, for example in a refrigeration system to record, was the 
compressor on, was the door open, was the lighting on, etcetera. External signal conditioning 
circuits are available to convert the external parameter to the levels necessary for the digital logic 
inputs:- 
 

Light sensor

, detects light levels and inputs a digital signal to the DS-108 when the light is above 

a set level. 

 

Mains voltage sensor

, detects 120/230Vac and inputs a digital signal to the DS-108 when 

voltage is present. 

 

Magnetic field sensor,

 detects magnetic flux density levels and inputs a digital signal to the DS-

108 when the flux is above a set level, this can be used to detect the proximity of an item or 
whether or current is flowing in a conductor. 

 

Other input conditioning circuits may become available, please contact your local supplier 
regarding any special needs you have. 

 
The voltage inputs and the +5V/400mA output are on the rear of the product. The 5V out is rated 
at 400mA maximum. Grounds (Gnd) is common and also common to the USB Gnd. The label on 
the underside of the logger detail the input and output connections. The inputs have a maximum 
of +30Vdc with respect to Gnd (0V). The 5Vdc output can be used to power conditioning circuitry 
or for any other use you may have, it is fused internally to 400mA. Logic inputs 0Vdc to ~2Vdc are 
considered 0 (off)  and ~2.1Vdc to 30Vdc logic 1 (on) 
 