background image

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

FZ-120

i

WP / FZ-200

i

WP / FZ-300

i

WP 

FZ-1200

i

WP / FZ-2000

i

WP / FZ-3000

i

WP 

 

FX-120

i

WP / FX-200

i

WP / FX-300

i

WP 

FX-1200

i

WP / FX-2000

i

WP / FX-3000

i

WP 

 
 
 
 
 
 
 
 
 
 
 
 

1WMPD4001591A 

Summary of Contents for FZ-120iWP

Page 1: ...FZ 120iWP FZ 200iWP FZ 300iWP FZ 1200iWP FZ 2000iWP FZ 3000iWP FX 120iWP FX 200iWP FX 300iWP FX 1200iWP FX 2000iWP FX 3000iWP 1WMPD4001591A...

Page 2: ...or translated into any language in any form by any means without the written permission of A D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual a...

Page 3: ...Counting Mode PCS 17 5 3 Percent Mode 19 5 4 Statistical Calculation Mode 20 5 5 Statistical Calculation Mode Example of Use 25 Adapting to the Environment 6 RESPONSE ADJUSTMENT 27 7 CALIBRATION 28 7...

Page 4: ...UNDERHOOK 53 12 PROGRAMMABLE UNIT 54 RS 232C Serial Interface 13 RS 232C SERIAL INTERFACE 55 14 CONNECTION TO PERIPHERAL EQUIPMENT 56 14 1 Connection to the AD 8121B Printer 56 14 2 Connection to a Co...

Page 5: ...ose balance can be installed almost anywhere Casing construction strong protection against dust and water Stabilization time of one second When FAST is selected for the response rate a stabilization t...

Page 6: ...interference and under these circumstances the user would be required to take at his own expense whatever measures are necessary to eliminate the interference FCC Federal Communications Commission in...

Page 7: ...5...

Page 8: ...6...

Page 9: ...able The breeze break is an accessory for FZ i WP series FX 120i WP FX 200i WP FX 300i WP The breeze break Assemble the breeze break as shown below Follow the numbered sequence 4 1 2 3 Waterproof diap...

Page 10: ...2000i WP FX 3000i WP Assemble the pan support and weighing pan on the balance as shown in the illustration on page 7 3 Adjust the leveling feet to level the balance Confirm it using the bubble spirit...

Page 11: ...y are not excessive The best operating temperature is about 20 C 68 F at about 50 relative humidity Install the balance where it is not exposed to direct sunlight and it is not affected by heaters or...

Page 12: ...ambient humidity above 45 RH Use a metal shield case for a charged sample Wipe charged plastic samples with a moistened cloth This balance uses a strong magnet as part of the balance assembly so pleas...

Page 13: ...spray 3 3 After Use Do not allow the balance to be immersed in water Even though the balance complies with IP code the balance will not withstand being completely immersed in water Avoid mechanical sh...

Page 14: ...n table mode Refer to 9 FUNCTION TABLE Switches the weighing units stored in the function table Refer to 4 WEIGHING UNITS Enters the response adjustment mode Cancels the operation when performing sett...

Page 15: ...ode for weighing press the MODE key For details about the units and modes see the table below Name unit mode Display Function table Storing mode Conversion factor 1 g Gram g g 1 g Counting mode pcs pc...

Page 16: ...l Taiwan 3 25330 5 86670 8 53330 0 00005 Tael China 3 90400 7 04000 10 24000 0 00005 Tola India 10 4597 18 8618 27 4353 0 0001 Messghal 26 0265 46 9335 68 2665 0 0005 Capacity Unit FX 1200i WP FX 2000...

Page 17: ...pecify a unit or mode in the order to be displayed using the following keys SAMPLE key To sequentially display the units RE ZERO key To specify a unit or mode The stabilization indicator appears when...

Page 18: ...While the stabilization indicator is on pressing the PRINT key will output the weight value using the RS 232C serial interface Note Peripheral equipment that is sold separately such as a printer or a...

Page 19: ...A greater number of samples will yield more accurate counting result 4 Place a container on the weighing pan if necessary Press the RE ZERO key to cancel the weight tare The number specified in step 3...

Page 20: ...unit mass automatically by increasing the number of samples and averaging the unit mass variable to minimize the weighing error as the counting process proceeds 8 If a few more samples are added the p...

Page 21: ...mass on the pan or in the container 5 Press the PRINT key to store the reference mass The balance displays 100 00 The decimal point position depends on the reference value The reference mass stored e...

Page 22: ...r the input Turning the balance off will delete the statistical data The standard deviation and coefficient of variation are obtained by the equation below Standard deviation where Xi is the i th weig...

Page 23: ...ficient of variation 7 Press the PRINT key to store the setting Note The statistical calculation mode will be enabled using the ON OFF key or when the power is turned on 8 Press the CAL key to return...

Page 24: ...a number and the weight data and includes the weight data to statistical calculation in the weighing mode Output is not in the data format specified in the function table because of the data number ad...

Page 25: ...and coefficient of variation are displayed as When the average is 0 the coefficient of variation is displayed as Statistical items are indicated on the upper left of the display using the following s...

Page 26: ...te the latest data and exclude it from statistical calculation The number of data decreases by 1 when the balance returns to the weighing mode Clearing the statistical data All the statistical data wi...

Page 27: ...n release the key 2 Select the application function Press the SAMPLE key several times to display ap fnc Then press the PRINT key to display apf 0 3 Change the application function parameter to 2 Pres...

Page 28: ...0 00 g Storing the weight value of formula 1 The weight value data is output when the peripheral output equipment is connected 4 Weigh formula 2 and press the PRINT key The balance displays 0 00 g St...

Page 29: ...SE is displayed And then release the key 2 Press the MODE key to select a rate of the response adjustment Either FAST MID or SLOW can be selected 3 After a few seconds of inactivity or when the PRINT...

Page 30: ...ation or drafts to affect the balance during calibration To output the data for GLP using the RS 232C serial interface set GLP output info of Data output dout For details refer to 9 FUNCTION TABLE Cal...

Page 31: ...n If the GLP output info parameter of the function table is set to 1 or 2 the balance displays glp and outputs Calibration Report using the RS 232C interface or stores the data in memory For details o...

Page 32: ...ibration weight value stored in the balance proceed to step 5 4 Specify the calibration weight value as follows SAMPLE key To switch the display condition to All of the segments blinking calibration w...

Page 33: ...ight Do not allow vibration or drafts to affect the balance 7 The balance displays end Remove the weight from the pan 8 When the GLP output is set the balance displays glp and outputs Calibration Repo...

Page 34: ...ou want to change the target weight a list of usable weights is shown on page 28 press the SAMPLE key and proceed to step 4 If you use the target weight value stored in the balance proceed to step 5 4...

Page 35: ...e the displayed target weight on the pan and press the PRINT key The balance measures the target weight and displays the measured value Do not allow vibration or drafts to affect the balance 7 The bal...

Page 36: ...hes FX iWP series Function table 0 To inhibit changes to the function table 1 To permit changes to the function table Calibration using the internal mass One touch calibration 0 To inhibit calibration...

Page 37: ...e CAL key once again 8 2 Initializing the Balance This function returns the following parameters to factory settings Calibration data Function table The sample unit mass value counting mode 100 refere...

Page 38: ...bol indicates that the parameter displayed is in effect When pressed and held in the weighing mode enters the function table mode Selects the class or item in the function table mode Changes the param...

Page 39: ...ma Decimal point format 0 OFF p on Auto display ON 1 ON Turns on the weighing mode display when the AC adapter is connected 0 OFF poff Auto display OFF 1 ON 10 minutes Turns off the display after 10 m...

Page 40: ...erence 1 Minus only Displayed value Reference ap p Auto print polarity 2 Both Regardless of displayed value 0 10 digits 1 100 digits ap b Auto print difference 2 1000 digits Difference between referen...

Page 41: ...able unit Multi unit Sets an arbitrary coefficient Available only when programmable unit mode is selected Unit Unit Refer to 4 WEIGHING UNITS id ID number Refer to 10 ID NUMBER AND GLP REPORT 0 Normal...

Page 42: ...t a weight value from drifting due to vibration or drafts Note With Hold function Hold set to ON 1 this item is used to set the stabilization range Hold function Hold Animal weighing mode This functio...

Page 43: ...rong zero tracking Display refresh rate 5pd Period to refresh the display This parameter influences Stream mode along with Baud rate and Data output pause Note This item is selected automatically when...

Page 44: ...added with ar d set to 1 to adjust zero after the data is output Mode B Required setting dout prt 2 Auto print mode B reference laststablevalue dout ap p Auto print polarity dout ap b Auto print diffe...

Page 45: ...at If an AD 8121B is used set the printer to MODE 1 or 2 This format consists of fifteen characters excluding the terminator A header of two characters indicates the balance condition The polarity sig...

Page 46: ...format 5if type 3 A header of two characters indicates the balance condition The polarity sign is used only for negative data The weight data uses spaces in place of the leading zeros The character le...

Page 47: ...45 9 7 Data Format Examples...

Page 48: ...pacity indicator apf 1 In the weighing mode the indicator displays the weight data relative to the weighing capacity in percentage Zero 0 maximum capacity 100 Statistical calculation mode apf 2 The mo...

Page 49: ...mode Entering the upper limit value 6 With Cp Hi displayed press the PRINT key The current setting of the upper limit value is displayed with all the digits blinking When the current setting is not t...

Page 50: ...digits blinking 5 Set the Year Month and Date using the following keys SAMPLE key To select the digits to change the value The selected digits blink RE ZERO key To increase the value by one MODE key...

Page 51: ...es Do not enter invalid values such as a non existing date when setting the time and date When the clock backup battery has been depleted the balance displays rtc pf Under this condition press any key...

Page 52: ...ock and End block for the weight data 10 1 Setting the ID Number 1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed 2 Press the SAMPLE key several times to display id 3 Pr...

Page 53: ...ABCDEFG DATE 2006 02 06 TIME 12 23 34 CALIBRATED EXT CAL WEIGHT 2000 00 g SIGNATURE TERM Space ASCII 20h Terminator or Carriage return ASCII 0Dh Line feed ASCII 0Ah CR CR LF CR LF AD 8121 format Gene...

Page 54: ...n the setting is info 2 Weight data Time Signature End block AD 8121 format Manufacturer Model Serial number ID number Date Time Title block General data format A D TERM MODEL FX 3000i TERM S N 012345...

Page 55: ...as shown below Caution Do not apply excessive force to the underhook When the underhook is not in use attach the plastic cap to prevent dust from getting into the balance The cap was sealed to the bot...

Page 56: ...the coefficient Confirming the coefficient 4 The current coefficient is displayed with the first digit blinking When it is not to be changed press the CAL key and proceed to step 6 When it is to be c...

Page 57: ...ta bits 7 bits None Data bits 8 bits Stop bit 1 bit Code ASCII D Sub 9 pin assignments Pin No Signal name Direction Description 1 No connection 2 TXD Output Transmit data 3 RXD Input Receive data 4 No...

Page 58: ...bp5 Baud rate 2 2 2 2 btpr Data bit parity bit 0 0 0 0 Crlf Terminator 0 0 0 0 5if Serial interface type Data format 0 0 0 1 1 Set appropriate parameters for ap p Auto print polarity and ap b Auto pri...

Page 59: ...it the balance data to a personal computer The FX iWP series balance is a DCE Use a straight through cable If purchasing the RS 232C cable on the market check the interface connections and type Use th...

Page 60: ...osoft Excel Can be used with most application software Using the WinCT software the balance can do the following 1 Analyzing the weight data and the statistics with RsKey The weight data can be input...

Page 61: ...en stabilized Note The Q and SI commands the S and E SCP commands behave the same Commands to control the balance CAL Same as the CAL key OFF Turns the display off ON Turns the display on P Same as th...

Page 62: ...ommand to control the balance and can not process it the balance transmits an error code EC Exx When the balance receives a command to control the balance and can process it the balance transmits the...

Page 63: ...oise or a parity error has occurred due to transmission error the balance transmits an error code In this case send the command again 15 3 Settings Related to the RS 232C Concerning the RS 232C the ba...

Page 64: ...D dealer if the balance needs service or repair Use the original packing material for transportation While cleaning the balance and keeping it waterproof cover the RS 232C interface connector using th...

Page 65: ...ance level Refer to 3 1 Before Use Is the operating environment free from vibration and drafts For the FZ 120i WP FZ 200i WP FZ 300i WP FX 120i WP FX 200i WP FX 300i WP has the breeze break been insta...

Page 66: ...to return to the weighing mode EC E21 Calibration weight error The calibration weight is too light Confirm that the weighing pan is properly installed Confirm the calibration weight value Press the C...

Page 67: ...one second Confirm the communication EC E04 Excess characters error The balance received excessive characters in a command Confirm the command EC E06 Format error A command includes incorrect data e g...

Page 68: ...as Microsoft Excel Word and memo pad Driver installation is not necessary FXi 08 Ethernet interface Used to connect the balance to a LAN The WinCT Plus data communication software is provided as an ac...

Page 69: ...more height AX USB 9P EX USB converter Adds a COM port to a PC Enables bi directional communication between the PC and the balance when a USB driver is installed Can use serial communication software...

Page 70: ...FX i WP series balance using the RS 232C serial interface to display the weight data away from the balance AD 8922 Remote controller Connected to the FX i WP series balance using the RS 232C serial i...

Page 71: ...pieces Minimum 100 reference mass 0 100 g 1 00 g Percent mode Minimum 100 display 0 01 0 1 1 Depends on the reference mass stored External calibration weight 100 g 50 g 200 g 100 g 300 g 200 g 100 g 1...

Page 72: ...rate 5 times second 10 times second or 20 times second Minimum unit mass 0 001 g 0 01 g Counting mode Number of samples 5 10 25 50 or 100 pieces Minimum 100 reference mass 0 100 g 1 00 g Percent mode...

Page 73: ...weighing pan up to the cap of the breeze break Inside dimension 5 Height of the opening when a clear plate is removed 6 Underhook position FZ 1200i WP FZ 2000i WP FZ 3000i WP FX 1200i WP FX 2000i WP F...

Page 74: ...72 FXi WP 11 Large breeze break Unit mm 199 5 112 Internal size above the pan Opening 294 6 156 When fully opened Opening 270 105 31 198 107 When fully opened When fully opened When fully opened...

Page 75: ...rational function Re zero To set the display to zero GLP Good Laboratory Practice Repeatability Variation in measured values obtained when the same weight is placed and removed repetitively Usually ex...

Page 76: ...arity 38 AVE Average 23 Average 20 23 39 AX KO2737 500 3 9 57 62 67 AX USB 9P 67 B ba5fnc Environment display 15 37 Baud rate 39 55 56 beep Beep 37 bep HI buzzer 37 bep LO buzzer 37 bep OK buzzer 37 b...

Page 77: ...Inhibit 34 Initializing the Balance 35 int Interval time 38 Interval output mode 38 43 Interval output mode active indicator 12 Interval output mode standby indicator 12 Interval time 38 IP65 3 9 53...

Page 78: ...dicator 12 15 Stabilization time 69 70 Standard deviation 20 23 39 Standby indicator 12 5taf Statistical function mode output items 21 39 Statistical calculation mode 3 20 21 39 46 Statistical functio...

Page 79: ......

Page 80: ...es Office Gro e Stra e 13 b 22926 Ahrensburg Deutschland Telefon 49 0 4102 459230 Telefax 49 0 4102 459231 A D MERCURY PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 830...

Reviews: