background image

 

 
 
Published:  May 2014 

 

Rev: 1 Replaces all previously published guides on this topic until superseded. 

 

The following is a Quick Start Guide for reference purposes only.  

Before using the TR-600 PAPR, all users must read 

and understand the product User Instructions.

  If you have questions, consult your supervisor or call 3M Technical 

Service at 1-800-243-4630 (USA) or 1-800-267-4414 (Canada).   
 

1

.  

Inspect the TR-600 PAPR system. 

 If parts are missing or damaged, replace only with TR-600 replacement parts before 

proceeding. 

 

Motor/blower

.  Inspect for cracks, holes, damage to user interface panel and belt attachments.  

 

Battery pack

. Confirm case is intact.  Press TEST button on pack to confirm charge is sufficient for duration of 

the work period. 

 

Breathing tube

. Examine breathing tube for tears, holes or cracks. Bend the tube to verify that it is flexible and not 

showing signs of wear. Ensure the o-rings located at both ends of the breathing tube are present and intact. 

 

Filter/cartridge

. Closely inspect filter/cartridge plastic housing including the corners and latches, outer 

rectangular barrier, and inner circular filter seal gasket for cracks, tears, cuts, distortion, indentations or debris. 
Replace filter/cartridge if damaged. If the filter/cartridge has been mishandled or dropped, re-inspect fully. If you 
have any concern, contact 3M Technical Service for guidance. 

 

Headgear.  

Inspect

 

the headgear for tears, cuts, loose stitching, cracks, discoloration or other signs of damage.   

 

2.

  

Assemble TR-600 PAPR system

 

 

Attach battery pack to bottom of motor/blower.  Pack latches with a distinct click. Gently pull to confirm a secure 
connection.  

 

(If being used) Install spark arrestor, then prefilter into filter cover. 

 

Place the filter/cartridge into the filter cover, ensuring bottom latching tab snaps into place. Note: If not using the 
filter cover, proceed to next step and attach the filter/cartridge directly to the motor/blower.  

 

Place the hinge side of the filter/cartridge into the motor/blower, then snap latch side into the filter latch. 
Filter/cartridge latches with a distinct click. Gently pull to confirm a secure connection.  

 

Attach motor/blower to belt or backpack.

 

 

Insert the end of the breathing tube with the two small prongs into slots in the motor/blower air outlet. Twist the 
breathing tube ¼ turn to the right (clockwise) to lock.

 

 

Attach breathing tube to headgear. Push the QRS end of the breathing tube (blue pinch clip) onto the air inlet of the 
headgear. Headgear will secure with a distinct click. Gently pull to confirm a secure connection.  

 

3.  Perform flow check.

   

 

If installed, remove breathing tube. 

 

Insert air flow indicator TR-971 into the outlet on the TR-600 motor/blower unit. 

 

Start motor/blower. Run for 1 minute.    

 

With the airflow indicator TR-971 in a vertical position, bottom of the floating ball must be at, or above, the 
minimum flow level indicated on airflow indicator chart for your ‘zone’

.  If not above minimum level, respirator 

system must not

 

Remove air flow indicator TR-971 and perform low flow alarm test per TR-600 

User Instructions

.

 

 be used until evaluated and repaired.

 

 

Reattach breathing tube and headgear before use.

 

 

4.

  

Don TR-600 PAPR system

 

 

Turn on motor/blower by pressing and holding blue on/off button. 

 

Use belt or backpack to attach PAPR to wearer.

 

 

Pull headgear over head and adjust. 

Adjust following the user instructions for the specific headgear

 

5.

  

Post Use

 – Exit work area. Clean and store PAPR following practices established by your employer and in compliance 

with the product 

User Instructions. 

 

 
 

Quick Start Guide  
3M™ Versaflo™ Powered Air Purifying 
Respirator TR-600 Assemblies 

Reviews: