Using the O2 CALI function
1.
The O2 CALI function is as same as CO2’s.
*Note: When do the calibration, When
do the “ZEr” Calibration, please enter through N2. When do the “SPA”
Calibration, please enter through 21%
O2 gas. Before doing “ZEr”/” SPA” calibration, it will take about 5 min to
enter through O2 to stabilize. When do
“ZEr”/” SPA” calibration, the reading is beating too often and calibration
fails.
Using the ReFactSet function
1.
Press Mode until the “ReFactSet” icon flashes.
2.
Press Enter, and then press Mode to choose either “Yes” or “No”.
3.
Press Enter again to save the setting after selecting.
*Note: If the user sets the data or calibrates the sensor incorrectly, use the ReFactSet (recover the factory
Setting) to come back the default factory setting.