C

HAPTER

 3 O

PERATION

3-8

www.zoll.com

 9650-000272-05 Rev. B

Testing the Battery Bay

In the 

SurePower Single Bay Charger

, the control panel’s LEDs can be tested to determine if 

they are working correctly.

To test the control panel’s LEDs, follow this procedure:

Step

Action

1.

If there is a battery pack in the charging bay, remove it.

2.

Press the

 Test

 Button on the battery bay’s control panel.

The charger light all four LEDs for several seconds, and then extinguish them.

Summary of Contents for SurePower Single Bay Charger

Page 1: ...SurePower Single Bay Charger Operator s Guide 9650 000272 05 Rev B...

Page 2: ...information updates are available SurePower and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the prope...

Page 3: ...ePower II 1 2 Battery Bay Panel 1 3 Charger and Battery Keys 1 4 Battery Charging Requirement 1 4 Charging Modes 1 5 QuickCharge 1 5 ManualTest 1 5 Chapter 2 Set Up Powering on the Charger 2 2 Self Te...

Page 4: ...nd SurePower II Battery Pack 5 1 Developing a Battery Management Program 5 1 Chapter 6 Product Specifications SurePower Single Bay Charger Appendix A Manufacturer s Declaration Electromagnetic Emissio...

Page 5: ...s SurePower Single Bay Charger which you use to test recalibrate and charge ZOLL s rechargeable lithium ion defibrillator batteries It is important that you read and understand the information in thi...

Page 6: ...r 3 Operation Describes how to use the SurePower Single Bay Charger to test recalibrate and charge ZOLL s rechargeable defibrillator batteries Chapter 4 Maintenance and Troubleshooting Describes how t...

Page 7: ...er to Operator s Guide for important safety related information such as Warnings and Cautions that cannot be presented on the device DANGER high voltage present Protective earth ground terminal Altern...

Page 8: ...face type Return to a collection site intended for waste electrical and electronic equipment WEEE Do not dispose of in unsorted trash With respect to electric shock fire mechanical and other specified...

Page 9: ...container is damaged also notify the carrier Contacting Technical Service The SurePower Single Bay Charger does not require periodic recalibration or adjustment You should however periodically perform...

Page 10: ...the SurePower Battery Pack is AG The first two characters of the date of manufacture code give the last two digits of the year for example 05 would appear for products manufactured in 2005 The last c...

Page 11: ...Use only the AC cord supplied with the device Failure to use the proper line cord could result in excess leakage currents EMC problems and reduced safety Do not disassemble the SurePower Single Bay C...

Page 12: ...Monitor Defibrillators or a SurePower Single Bay Charger that is not plugged into live AC mains when storage may exceed 90 days Battery damage may occur Do not connect any cables or devices to the co...

Page 13: ...lmsford MA 01824 4105 Fax 978 421 0025 Tel 978 421 9655 Notification of Adverse Events As a health care provider you may have responsibilities under the Safe Medical Devices Act SMDA for reporting the...

Page 14: ...This page intentionally left blank...

Page 15: ...ingle Bay Charger is a single bay unit that can test recalibrate and charge a single ZOLL rechargeable defibrillator battery Figure 1 1 SurePower Single Bay Charger The charging bay in the SurePower S...

Page 16: ...arger with SurePower Battery Pack The following illustration is of a SurePower Single Bay Charger with a SurePower Battery Pack installed in it Charger with SurePower II Battery Pack The following ill...

Page 17: ...to indicate that the battery is charging Ready The LED lights next to this icon to indicate that the battery is fully charged and ready to use Test button Pressing this button initiates a manual test...

Page 18: ...ion shows the key on the SurePower Single Bay Charger Figure 1 3 Battery Key Batteries that do not have the correct keys will not seat in the charging bay and cannot be charged and tested by the SureP...

Page 19: ...harging method for ZOLL s lithium ion batteries such as the SurePower Battery Pack The QuickCharge mode is the quickest way to charge lithium ion batteries which do not require frequent testing to mai...

Page 20: ...This page intentionally left blank...

Page 21: ...ion Figure 2 1 SurePower Single Bay Charger Setup Caution Do not place anything on top of or beneath the unit that might block the vents Do not remove or discard the charger s port connector plug remo...

Page 22: ...power is available Position the SurePower Single Bay Charger so that the AC cord is easy to disconnect Warning Always inspect the AC cord before use Do not use the AC cord if it is defective for examp...

Page 23: ...SurePower Single Bay Charger Operator s Guide 3 1 Chapter 3 Operation The SurePower Single Bay Charger is designed to charge and recalibrate two different types of batteries the SurePower and SurePow...

Page 24: ...ce the back end of the battery pack against the back of the charging bay where the pins are not located and then press the battery down firmly until it clicks Warning To avoid pinching your fingers ho...

Page 25: ...te after you insert the battery 3 Allow the battery time to charge The charging time will depend on the battery type how depleted the battery was when you put it into the charger and which charging me...

Page 26: ...lip to disengage it from the latch and then lift the battery out of the charging bay Note If the Fault LED is on when you remove the battery pack the charger beeps to alert you that there has detected...

Page 27: ...clean undamaged charger 3 If the battery pack and charger are clean and undamaged reinsert the battery pack 4 If the Fault LED lights after reinserting the battery pack remove the battery pack wait 1...

Page 28: ...Amphere hours discharged You can also initiate a manual battery test which recalibrates the battery and measures charge capacity after a shorter service interval for example after a specific number of...

Page 29: ...the bay s control panel The charger lights the Test LED and begins testing the battery The Test LED will remain lit for the duration of the test and the Charging LED during the charging portion of th...

Page 30: ...anel s LEDs can be tested to determine if they are working correctly To test the control panel s LEDs follow this procedure Step Action 1 If there is a battery pack in the charging bay remove it 2 Pre...

Page 31: ...what the LED and alarm activity indicates during different operating modes The charger light LEDs in one of two ways Steady persistent or brief periodic Indicators When Powered On or Testing the Unit...

Page 32: ...nual Test Method battery charging On steady On steady Battery Capacity Test successful Manual Test Method On steady On steady Battery Capacity Test successful Automatic Recalibration On steady On stea...

Page 33: ...lators and monitors Disabling Enabling the Audible Alarm You can disable and enable the charger s audible alarm To disable or enable the audible alarm press and hold down the charger s Test button for...

Page 34: ...This page intentionally left blank...

Page 35: ...eep turn on the fan and light all LEDs for several seconds The Power LED will remain lit after the other LEDs go off 4 With the battery bay empty check that all LEDs on the control panel except the Po...

Page 36: ...ed in the bay Check that the contacts are clean and not damaged Use another battery that you know is good 3 Test LED does not light when you press the Test button Check that the battery is properly se...

Page 37: ...ePower or SurePower II Battery Packs in a fully depleted state this will damage the battery Never store the SurePower or SurePower II Battery Packs in a charger that is not powered on Developing a Bat...

Page 38: ...cardiopulmonary resuscitation should a battery pack or resuscitation device fail and need to be replaced 6 Test Battery Packs regularly Develop a testing schedule as part of your organization s Batte...

Page 39: ...Current 0 to 2 5 A dc Maximum Thermal Output 70 BTU HR Design Standards Designed to meet or exceed IEC 60601 1 standards for medical equipment safety Temperature Operation 32 F to 122 F 0 C to 50 C S...

Page 40: ...Maximum Charge Time SurePower and SurePower II Batteries 4 5 hours at 25 C 77 F 8 5 hours at 50 C 122 F Battery Contact Pins Usage Limit 5 000 removal insertion cycles...

Page 41: ...such an environment RF emissions CISPR 11 Group 1 The SurePower Single Bay Charger does not contain any intentional RF transmitters Therefore its RF emissions are very low and are not likely to cause...

Page 42: ...e that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip i...

Page 43: ...ion distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference ma...

Page 44: ...unications equipment Rated Maximum Output Power of Transmitter W Separation Distance According to Frequency of Transmitter m 150 kHz to 80 MHz d 1 2 80 MHz to 800 MHz d 0 35 800 MHz to 2 5 GHz d 0 7 0...

Reviews: