MONITOR MODE (

Grey Zone

)

 

LEAD 

- Change pressing lead. Circulates through Lead I-III and PADS 

SIZE 

- Change to make the waveform tracing bigger or smaller 

ALARM SUSPEND 

 

Default alarm volume is Off

 

 

Alarms On 

– Press Alarm Suspend button to turn the alarms on

 

 

Temporary Silence - 

Press again and it will be temporarily disabled the alarms (15 sec for lethal 

and 90 sec for non-lethal rhythms)

 

 

Alarms Off - 

Press and hold for 3

-

4 seconds and the alarms will be SILENCED until otherwise 

activated by the user. (an x will appear next to alarm icons)

 

  If the patient is in VFib, Vtac, or Asystole, the alarms will automatically sound after 5 seconds.

 

RECORDER 

- Starts or stops the paper recorder at any time 

 

SOFT KEYS

 

NIBP 

(the button with the picture of an arm, bottom left) 

  Press once for one (1) NIBP reading. Press again to abort reading.

 

  Press and hold for 2 Seconds for 

STAT 

mode (as many NIBP measurements as 

possible in 5 minutes)

 

Options 

 

QRS Vol On 

- Turns tone on or off

 

 

Low Bright 

- Adjust Screen Brightness

 

 

Traces 

- Waveforms you can bring up. (Options——Traces——Trace 2,3)

 

 

More 

- Reveals 

“Clock Set” 

to change time on display

 

Param 

 

EtCO2 Setup 

(

ONLY DONE ONCE 

when a new sensor is connected to this R Series)

 

1.  Insert EtCO2 Single Patient Airway Adapter 
2.  Press the “Param” Key 
3.  Press the “EtCO2” Key. 
4.  Press the “Zero” key. This will take about 10 seconds and will 

display “Zeroed” when done. 

 

NIBP Setup

 

1.  Press the “Param” Key 
2.  Press the “Setting” Key 
3.  Press Auto Interval 

4. 

Set time for Auto Interval (default 5 minutes)

 

Code Marker 

 

Varies with where the dial is positioned (Defib, Pacer, Monitor)

 

 

Allows you to time stamp different clinical actions and retrieve later on

 

Report Data 

(Printing and Data Transfer) 

 

PRINT

 

o

 

Print Chart: 

ECG strips for a specific time or all activated events 

o

 

Print Range: 

Select and print specific event 

o

 

Print Log

: Chronological list of all interventions since the device has been on for the 

current code and previous codes that are stored 

 

TEST LOG 

– Log of up to the last 1000 tests performed.

 

 

TRANSFER MODE 

– Print to WiFi

 

Alarms 

  Set highs and lows for different parameters.

 

  The alarms will be defaulted back once the device is off for 10 Seconds.

 

Reviews: