Connecting and Disconnecting 

  Locking lever holds pads and cable in place. Press on the 

far end of the lever to release.

 

 

Placement 

 

Do good skin prep! 

– Remove sweat, lotions, excess 

hair.

 

 

Posterior Pad 

(

RED

)

 

o

  Goes on first 

o

  Placed sub scapular to the left 

 

Anterior Pad 

(

BLUE

)

 

o

 

CPR sensor Mid Sternal/Mid Nipple 

center of the 

sternum (where you do compressions) 

o

  Pad should be off-right of the sensor. 

 

Press and Roll” 

the pad into place

 

Cables/Caps 

1.  Electrotherapy Cable (Is bolted down) 
2.  SpO2 (Blue/If Applicable) 
3.  3 Lead EKG and/or the cap in place for pacing 
4.  EtCO2 (Yellow) 
5.  NIBP- Non-Invasive Blood Pressure (Black) 
6.  Cable Compartment for all cables attached 

 

One-Step Cable: 

Comes out the side and is 8 feet in length. 

 

ONESTEP PAD

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Alternate Placement 

(If A/P is not applicable) 

 

Posterior Pad (

RED

o

  Placed on upper right chest 

 

Anterior Pad (

BLUE

o

  Placed on left anterior-lateral, remove CPR sensor and place 

in center of chest for compressions. 

 

Add 3-lead 

 

Pediatric Pads 

  Smaller and more suitable. 

  Any defibrillation that is done will be adjusted by the device. (50J, 70J, 85J) 

 

Integrated 3-Lead ECG  

 

  OneStep Adult pads have a built in 3-lead for pacing patients without an extra 3 lead cable. 

  If attempting to pace with alternate OneStep pad placement (A/A), it is recommended to 

attach a separate 3-Lead. Inverting the pad will invert the ECG. 