Page 8

Monitoring (Cont.)

SpO2

1. Connect the SpO2 cable to the patient sensor and to  

the blue SpO2 port on the rear of the unit labeled 

SpO2

.

2. Attach sensor to patient’s finger.
3. Turn selector switch to grey area labeled MONITOR on 

the R Series ALS. If using BLS or Plus device, turn ON, 
then press 

Manual

 then 

Confirm

 softkeys to use  

in ALS mode.

4. If properly attached, the SpO2 values will display  

on the screen.

5. To display waveform: Press 

Options

 softkey.  

Press 

Traces

 softkey. Press 

Trace 2

 or 

Trace 3

 

softkey. Select SpO2 softkey. 

NIBP

1. Plug NIBP cable into the back of the unit in silver port marked NIBP.
2. Attach appropriate BP cuff to patient’s arm or thigh.
3. Turn selector switch to grey area labeled MONITOR. If using BLS or Plus device,  

press 

Manual

 then 

Confirm

 softkeys to use in ALS mode.

4. For ALS: Press the 

Param

 softkey, then press the NIBP soft key.

5. On BLS or Plus device, press illuminated blood pressure icon on  

bottom left to take measurement.

Alarms

1. Press the 

Alarms

 softkey to select alarms.

2. Press 

Return

 softkey to activate alarms.

3. To mute for 90 seconds:  Press ALARM SUSPEND button. 
4. To reactivate muted alarm: Press ALARM SUSPEND button.
5. To disable: Press and hold ALARM SUSPEND for 3 seconds. 

Note: Alarms are still active, but silent.

Summary of Contents for R Series

Page 1: ...Corporation in the United States and or other countries All other trademarks are the property of their respective owners 9650 1001 01 Rev E ZOLL Medical Corporation Worldwide Headquarters 269 Mill Ro...

Page 2: ...R Series Quick Reference Guide...

Page 3: ......

Page 4: ...ge 2 Details for applying OneStep Electrodes ECG cables and paddles with the R Series Adult Defibrillation Page 3 Step by step defibrillation for the R Series ALS and BLS Plus models CPR Page 5 Real C...

Page 5: ...e sternum paddle to the right of sternum just below clavicle 5 Place apex paddle left of nipple on anterior axillary line 6 Press paddles firmly against patient s skin BACK FRONT OneStep Complete Resu...

Page 6: ...nd follow the instructions 5 If rhythm is shockable defibrillator will charge and red SHOCK button will illuminate 6 If shock button illuminates shout Clear once all are clear press SHOCK 7 If no shoc...

Page 7: ...kable press the CHARGE button 5 Shout Clear and verify others are clear 6 Press SHOCK button Pediatric Defibrillation Follow the instructions as outlined above for your unit with the following excepti...

Page 8: ...a 100 bpm metronome will be heard until rate is correct The indicator word Rate will then disappear from the screen 7 If your R Series does not display a PPI your defibrillator has only basic CPR fee...

Page 9: ...appears then increase by 10 6 Check pulse to verify mechanical capture 7 To check patient s underlying rhythm press and hold 4 1 button Note When using BLS or Plus models manual mode must be activated...

Page 10: ...able 3 Remove all clothing covering the patient s chest Dry chest if necessary Clip or shave excessive chest hair to ensure proper adhesion of the electrodes 4 Snap leads onto each electrode 5 Peel of...

Page 11: ...y NIBP 1 Plug NIBP cable into the back of the unit in silver port marked NIBP 2 Attach appropriate BP cuff to patient s arm or thigh 3 Turn selector switch to grey area labeled MONITOR If using BLS or...

Page 12: ...closure Files to a Compact Flash Card or by WiFi 1 Press the Report Data softkey 2 Press the Transfer Mode softkey 3 Press the Report To Card softkey to transfer to a Compact Flash Card or press the R...

Reviews: