Monitoring

End Tidal CO2 (EtCO2) 

 Mainstream

1. Plug EtCO2 cable into the back of the unit in the yellow port marked 

CO2

.

2. Attach sensor to patient’s ET tube.
3. Turn selector switch to grey area labeled MONITOR. If using BLS or Plus device, 

turn ON, then press 

Manual

 then 

Confirm

 softkeys to use in ALS mode.

4. Press the 

Param

 softkey.

5. Press the CO2 softkey.
6. Press the 

Enable EtCO2

 softkey.

7. If properly attached, the EtCO2 values will display on the screen.
8. To display waveform: Press 

Options

 softkey. Press 

Traces

 softkey.  

Press 

Trace 2 

or 

Trace 3

 softkey. Select CO2 softkey.

ECG Monitoring

Using OneStep ECG Cable

1. Disconnect OneStep electrode from the OneStep therapy cable.
2. Plug the OneStep ECG cable into the OneStep therapy cable.
3. Remove all clothing covering the patient’s chest. Dry chest if necessary. Clip  

or shave excessive chest hair to ensure proper adhesion of the electrodes.

4. Snap leads onto each electrode.
5. Peel off protective backing from electrode.
6. Apply electrodes to patient’s skin.

Using Standard ECG cable

1. Unplug OneStep ECG cable from back of unit.
2. Plug ECG cable connector into back of unit.
3. Remove all clothing covering the patient’s chest.  

Dry chest if necessary. Clip or shave excessive chest 
hair to ensure proper adhesion of the electrodes.

4. Snap leads onto each electrode.
5. Peel off protective backing from electrode.
6. Apply electrodes to patient’s skin.

ECG Display 

1. Turn selector switch to MONITOR on the R Series ALS unit. If using BLS or Plus 

device, press 

Manual

 then 

Confirm

 softkeys to activate ALS mode.

2. Press LEAD button to change leads.
3. Press SIZE button to change size of display.

Dial settings for  

R Series Monitoring

Summary of Contents for R Series

Page 1: ...Corporation in the United States and or other countries All other trademarks are the property of their respective owners 9650 1001 01 Rev E ZOLL Medical Corporation Worldwide Headquarters 269 Mill Ro...

Page 2: ...R Series Quick Reference Guide...

Page 3: ......

Page 4: ...ge 2 Details for applying OneStep Electrodes ECG cables and paddles with the R Series Adult Defibrillation Page 3 Step by step defibrillation for the R Series ALS and BLS Plus models CPR Page 5 Real C...

Page 5: ...e sternum paddle to the right of sternum just below clavicle 5 Place apex paddle left of nipple on anterior axillary line 6 Press paddles firmly against patient s skin BACK FRONT OneStep Complete Resu...

Page 6: ...nd follow the instructions 5 If rhythm is shockable defibrillator will charge and red SHOCK button will illuminate 6 If shock button illuminates shout Clear once all are clear press SHOCK 7 If no shoc...

Page 7: ...kable press the CHARGE button 5 Shout Clear and verify others are clear 6 Press SHOCK button Pediatric Defibrillation Follow the instructions as outlined above for your unit with the following excepti...

Page 8: ...a 100 bpm metronome will be heard until rate is correct The indicator word Rate will then disappear from the screen 7 If your R Series does not display a PPI your defibrillator has only basic CPR fee...

Page 9: ...appears then increase by 10 6 Check pulse to verify mechanical capture 7 To check patient s underlying rhythm press and hold 4 1 button Note When using BLS or Plus models manual mode must be activated...

Page 10: ...able 3 Remove all clothing covering the patient s chest Dry chest if necessary Clip or shave excessive chest hair to ensure proper adhesion of the electrodes 4 Snap leads onto each electrode 5 Peel of...

Page 11: ...y NIBP 1 Plug NIBP cable into the back of the unit in silver port marked NIBP 2 Attach appropriate BP cuff to patient s arm or thigh 3 Turn selector switch to grey area labeled MONITOR If using BLS or...

Page 12: ...closure Files to a Compact Flash Card or by WiFi 1 Press the Report Data softkey 2 Press the Transfer Mode softkey 3 Press the Report To Card softkey to transfer to a Compact Flash Card or press the R...

Reviews: