Maintenance Tests
9650-0807-01 Rev. C
ZOLL Propaq MD Service Manual
43
24.0 NIBP Test
The primary purpose of an NIBP simulator is to reproduce a pressure profile similar to a live
patient to be used for testing repeatability and functionality of the system. There are many
different NIBP simulators, and manufacturers use different methods to develop their algorithm,
which can cause varying readings from different simulators. In order to test for repeatability,
you should first establish the offset of your simulator. The offset is a value that is produced on
the NIBP monitor that is shifted from the simulator’s setting. This value must be established
based on the a statistical sample of monitors and readings.
Note:
Refer to Appendix B for information on simulators and settings that you may use to
assess the performance of the NIBP module.
Equipment
Dual lumen NIBP hose, NIBP simulator
Do This...
Observe This...
Pass/Fail
24.1 On the Propaq unit, ensure the patient
mode is set to Adult.
24.2 Set the simulator to the 80/50 simulator
preset.
24.3 Take a manual reading.
Confirm that the NIBP measurement
reported is /- 10 mmHg of
the expected simulator performance.
o
o
24.4 Set the simulator to the 120/80 simulator
preset.
24.5 Take a manual reading.
Confirm that the NIBP measurement
reported is /- 10 mmHg of
the expected simulator performance.
o
o
24.6 Set the simulator to the 200/150 simulator
preset.
24.7 Take a manual reading.
Confirm that the NIBP measurement
reported is /- 10 mmHg of
the expected simulator performance.
o
o
24.8 Disconnect hose.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Service Manual 9650 0807 01 Rev C...
Page 6: ...TABLE OF CONTENTS iv www zoll com 9650 0807 01 Rev C...
Page 64: ...58 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 102: ...96 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 108: ...CHAPTER 4 REPLACEMENT PARTS 102 9650 0807 01 Rev C Pictures and Diagrams Picture 1 Picture 2...
Page 109: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 103 Picture 3 Picture 4...
Page 110: ...CHAPTER 4 REPLACEMENT PARTS 104 9650 0807 01 Rev C Picture 5 Picture 6...
Page 111: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 105 Picture 7 Picture 8...
Page 112: ...CHAPTER 4 REPLACEMENT PARTS 106 9650 0807 01 Rev C Picture 9 Picture 10...
Page 113: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 107 Picture 11 Picture 12...
Page 114: ...CHAPTER 4 REPLACEMENT PARTS 108 9650 0807 01 Rev C Picture 13 Picture 14...
Page 115: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 109 Picture 15 Picture 16...
Page 116: ...CHAPTER 4 REPLACEMENT PARTS 110 9650 0807 01 Rev C Picture 17 Picture 18...
Page 117: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 111 Picture 19 Picture 20...
Page 118: ...CHAPTER 4 REPLACEMENT PARTS 112 9650 0807 01 Rev C Picture 21 Picture 22...
Page 119: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 113 Picture 23 Picture 24...
Page 120: ...CHAPTER 4 REPLACEMENT PARTS 114 9650 0807 01 Rev C Picture 25 Picture 26...
Page 121: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 115 Picture 27 Picture 28...
Page 122: ...CHAPTER 4 REPLACEMENT PARTS 116 9650 0807 01 Rev C Picture 29 Picture 30...
Page 123: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 117 Picture 31 Picture 32...
Page 124: ...9650 0807 01 Rev C Propaq MD Service Manual 118...
Page 134: ...128 9650 0807 01 Rev C This page intentionally left blank...
Page 138: ...132 9650 0807 01 Rev C This page intentionally left blank...
Page 142: ...This page intentionally left blank...