ZOLL M Series Operator'S Manual Download Page 7

End TIdal Carbon Dioxide (EtCO

2

)

EtCO

2

 - 5

 

EtCO

2

 Indications for Use

The ZOLL M Series EtCO

option is indicated for the 

continuous noninvasive monitoring of end tidal carbon 
dioxide (EtCO

2

) and respiration rate in patients requiring 

ventilator support, transport, or anesthesia. The M Series

 

EtCO

2

 

option with Respironics Novametrix technology 

supports two methods for continuous measurement of 
end tidal carbon dioxide (EtCO

2

) and respiration rate. 

The first method uses the CAPNOSTAT 3 Mainstream 
CO

Sensor attached to an airway adapter that connects 

to an endotracheal tube, mask or disposable 
mouthpiece. 
The second method uses the LoFlo Sidestream CO

Module to monitor both non-intubated and intubated 
patients using specially designed sampling cannulae and 
airway adapters. 
The M Series

 

EtCO

2

 

option is designed to monitor adult, 

pediatric, and neonatal patients. 
The following substances can influence CO

measurements made with the CAPNOSTAT 3 Sensor:

elevated oxygen levels

nitrous oxide

halogenated agents

The M Series EtCO

2

 

option allows the user to enable 

high oxygen and/or nitrous oxide compensation. 
Halogenated anesthetic agents alter CO

readings, but 

the M Series unit will monitor CO

within specifications 

when these agents are present at normal clinical levels. 
The presence of Desflurane in the exhaled breath 
beyond normal values (5-6%) may positively bias 
measured carbon dioxide values by up to an additional 
2-3 mmHg. 
The M Series

 

EtCO

2

 

option is intended for use only with 

the ZOLL/Respironics Novametrix CAPNOSTAT 3 
Mainstream CO

sensor and the LoFlo Sidestream CO

Module, mainstream airway adapters, nasal and nasal/
oral sampling cannulae, and sidestream on-airway 
adapters.

Mainstream EtCO

2

 Setup

There are several steps involved with mainstream 
EtCO2 setup, as follows:

Attaching the CAPNOSTAT 3 sensor cable.

Selecting a mainstream airway adapter.

Attaching the airway adapter to the CAPNOSTAT 3 
sensor.

Zeroing the mainstream sensor/airway adapter.

Attaching the airway adapter to the airway circuit.

Applying an airway adapter with mouthpiece.

Attaching the CAPNOSTAT 3 Sensor Cable 

To attach the CAPNOSTAT 3 sensor cable, plug the 
cable into the CO

2

 connector at the back of the M Series 

unit.

Note

Effective with M Series software revision 38.35 and 
M Series CCT software revision 57.50, the “0” and 
“REF” cells on the sensor cable are no longer used. 
However, they remain on the sensor cable for 
backwards compatibility with earlier versions of the 
software.

Selecting a Mainstream Airway Adapter

Select an airway adapter based on the patient's size, ET 
tube diameter and monitoring situation. For more 
information refer to the following table or contact ZOLL 
Medical Corporation.

*SPU = Single Patient Use

Attaching the Airway Adapter to the 
CAPNOSTAT 3 Sensor

Before attaching the airway adapter to the 
CAPNOSTAT 3 sensor, verify that the airway adapter 
windows are clean and dry. Clean or replace the adapter 
if necessary.

CAUTION!

The disposable (SPU) Pediatric/Adult and 
the Neonatal/Pediatric airway adapters are 
intended for single patient use. Do NOT 
reuse or sterilize these adapters as system 
performance will be compromised.

1. Attach the airway adapter to the CAPNOSTAT 3 

sensor, as follows:

a. Align the arrow on the bottom of the airway adapter 

with the arrow on the bottom of the CAPNOSTAT 3.

b. Press the sensor and airway adapter together until 

they click.

2. If the unit displays the “CHECK CO2 ADAPTER” 

message, follow steps a through c, or else go to 
step 3.

a. Verify proper connection of the adapter to the 

sensor.

b. Verify that the airway adapter windows are clean 

and dry.

Airway Adapter Type

ET Tube Diameter

SPU Pediatric/Adult

> 4.0 mm

Adult Reusable

> 4.0 mm

SPU Neonatal/Pediatric

4.0 mm

Neonatal Reusable

4.0 mm

Summary of Contents for M Series

Page 1: ...9650 0223 01 Rev D TM End Tidal Carbon Dioxide EtCO2...

Page 2: ...l Corporation CAPNOSTAT is a registered trademark and LoFlo is a trademark of Respironics Novametrix LLC Cidex is a registered trademark of Advanced Sterilization Products a Johnson and Johnson Compan...

Page 3: ...mum filtration of fluids and contaminants and protecting the system from aspiration of these fluids In both systems the CO2 sensor generates infrared light and beams it through the airway adapter or s...

Page 4: ...es EtCO2 option Activate oxygen compensation if O2 levels in excess of 60 are introduced Activate N2O compensation if nitrous oxide is introduced into the airway circuit The presence of Desflurane bey...

Page 5: ...h windows in a vertical NOT a horizontal position This helps keep patient secretions from pooling on the windows Do NOT insert any object other than the sample cell into the sample cell receptacle on...

Page 6: ...t and property even if ZOLL has been advised of the possibility of such damages THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL MEDICAL CORPORATION EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES WHETHER...

Page 7: ...EtCO2 setup as follows Attaching the CAPNOSTAT 3 sensor cable Selecting a mainstream airway adapter Attaching the airway adapter to the CAPNOSTAT 3 sensor Zeroing the mainstream sensor airway adapter...

Page 8: ...ximately 15 seconds The unit displays the message ZERO DONE upon completion of the zeroing Note Do not attempt zeroing for 20 seconds after removing the adapter from the patient s airway This time all...

Page 9: ...ow steps a through c else go to step 4 a Verify proper connection of the adapter to the sensor b Verify that the airway adapter windows are clean and dry c If the adapter is properly connected and the...

Page 10: ...after its use When the previously used module is reconnected to the M Series unit module zeroing does not have to be repeated as long as no other LoFlo module has been connected and zeroed during the...

Page 11: ...lo module and ensure that it clicks into place 2 If the unit displays either of the following messages take the appropriate action 3 Place the airway adapter assembly at the proximal end of the airway...

Page 12: ...o module and rubber protective covering can be cleaned and disinfected by wiping with one or more of the following solutions 70 isopropyl alcohol 2 gluteraldehyde 10 bleach solution Use a soft cloth d...

Page 13: ...ftkey from the physiological monitoring menu cycles the display from the capnogram waveform to the plethysmograph waveform if SpO2 is installed to no second waveform displayed Use the Zoom softkey fro...

Page 14: ...airway circuit If the concentration of oxygen in the breathing circuit exceeds 60 and nitrous oxide is in use both O2 and N2O should be activated When the Comp softkey is pressed the unit displays the...

Page 15: ...alarm functions are disabled Background ECG analysis functions continue to operate as described in the AED section of the M Series Operator s Guide The ALARM SUSPEND button can be used to activate dea...

Page 16: ...ature unit displays WARM UP message 4 Perform a zero procedure if necessary see Zeroing the LoFlo CO2 Module on page 8 5 Breath normally into the adapter 6 Verify that the unit displays EtCO2 readings...

Page 17: ...3 kPa 0 95 mmHg OFF 0 12 5 OFF 0 12 6 kPa OFF High Respiration Rate Alarm Limit 120 respirations per min 5 150 respirations per min OFF Low Respiration Rate Alarm Limit 5 respirations per min 0 100 re...

Page 18: ...PU Oral Nasal CO2 sampling only pediatric Nasal CO2 Sampling with O2 Delivery Cannula Adult 8000 0356 SPU Nasal CO2 sampling with O2 delivery adult Nasal CO2 Sampling with O2 Delivery Cannula Pediatri...

Page 19: ...problem persists return for service CO2 DEVICE NOT READY The zero operation cannot be initiated because The sensor or module is still warming up No sensor or module is attached to the unit Wait for se...

Page 20: ...g the adapter type Clean airway adapter and reattach it If the problem persists or the adapter type was changed perform a mainstream airway adapter zero as described in Zeroing the Mainstream Sensor A...

Page 21: ...line is set below 0 mmHg Perform a sidestream module zero as described in Zeroing the LoFlo CO2 Module on page 8 CHECK CO2 LINE Sample line blockage Verify that the sample line is plugged into the mod...

Page 22: ...han 5 minutes Replace module ZERO CO2 MODULE Module cable connected to M Series unit for the first time Probe drift error detected or negative CO2 detected Perform module zero as described in Zeroing...

Page 23: ...lt pediatric adapters 50 ms Neonatal adapters 200 ms Respiration Rate Range 0 150 breaths per minute Respiration Rate Accuracy 1 breath per minute Sample Flow Rate N A 50 ml min Compensations Barometr...

Page 24: ...at maximum energy 360J or 1 5 hours minimum of continuous ECG monitoring or 1 0 hour of continuous ECG monitoring pacing at 60 mA 70 beats min For a new fully charged XL battery pack at 20 C 60 defibr...

Reviews: