ZOLL M Series Operator'S Manual Download Page 11

End TIdal Carbon Dioxide (EtCO

2

)

EtCO

2

 - 9

 

2. Press the 

Param.

 softkey and select the EtCO2 

menu item, then press 

Enter

3. Press the 

Zero

 softkey until the 

Start

 menu item is 

highlighted, then press 

Enter

The unit zeros the module and displays the 
"ZEROING CO2 MODULE" message for 
approximately 15 seconds.

The units displays the message “ZERO DONE” upon 
completion of the zeroing.

Note

Do not attempt zeroing for 20 seconds after 
removing the adapter or cannula from the patient’s 
airway. This time allows any CO

2

 

remaining in the 

adapter or cannula to dissipate before zeroing. Do 
not attempt to zero the module while the adapter or 
cannula is in the patient’s airway. Zeroing with CO

in the adapter or cannula can lead to inaccurate 
measurements and/or other error conditions. If you 
attempt zeroing while CO

remains in the adapter 

or cannula, the time required to zero the module 
may be increased. If zeroing cannot be completed, 
the message “ZERO FAILED” will be displayed. If 
this occurs, clear any occlusion in the adapter or 
cannula, remove the source of CO

2

, wait 20 

seconds, and try zeroing again.

Applying a Sidestream Airway Adapter Kit

The sidestream airway adapter kit is intended for 
monitoring the EtCO

2

 

of intubated patients.

Before attaching the airway adapter to the breathing 
circuit, verify that the adapter is clean, dry, and 
undamaged. Replace if necessary.

CAUTION!

The disposable (SPU) Adult/Pediatric and 
Pediatric/Infant airway adapter kits are 
intended for single patient use. Do NOT 
reuse or sterilize these adapters as system 
performance will be compromised.

1. Attach the airway adapter kit’s sample cell to the 

sample cell receptacle on the LoFlo module, and 
ensure that it clicks into place.

2. If the unit displays either of the following messages 

take the appropriate action.

3. Place the airway adapter assembly at the proximal 

end of the airway circuit between the elbow and the 
ventilator circuit wye. Do NOT place the airway 
adapter between the ET tube and the elbow as this 
may allow patient secretions to accumulate in the 
adapter. 

If pooling does occur, the airway adapter may be 
removed from the circuit, rinsed with water and 
reinserted into the circuit. To prevent moisture from 
draining into the sample tubing, ensure that the 
sampling tube exits from the top of the airway 
adapter, not its bottom or sides. See Figure 5.

4. Check that connections have been made correctly by 

verifying the presence of a proper capnogram on the 
M Series display.

Applying a Nasal or Nasal/Oral Cannula 

The nasal and nasal/oral cannulae are intended for 
monitoring EtCO

2

 in non-intubated patients. 

Oral/nasal sampling cannulae should be used on 
patients who are prone to mouth breathing, since most (if 
not all) of the CO

is exhaled through the mouth. If a 

If you see this 

message:

Take this action:

CHECK CO2 
LINE

Verify that the sample cell is plugged 
into the module and seated properly.
Verify that neither the sample line nor 
the exhaust tube are blocked, kinked, 
or pinched. Verify that the airway 
adapter is not blocked. If the sample 
line, exhaust tube, or the airway 
adapter is blocked or otherwise 
compromised, the pump shuts off in 
approximately 15 seconds. To restart 
the pump, correct the blockage, then 
remove and reinsert the sample cell 
into the sample cell receptacle.
If the problem persists, replace the 
sample line.

CHECK CO2 
MODULE

Check that module cable is plugged in 
and seated properly.
Check that module is not exposed to 
excessive heat.
If problem persists, replace module.

Figure 5

 

Sampling Tube

Summary of Contents for M Series

Page 1: ...9650 0223 01 Rev D TM End Tidal Carbon Dioxide EtCO2...

Page 2: ...l Corporation CAPNOSTAT is a registered trademark and LoFlo is a trademark of Respironics Novametrix LLC Cidex is a registered trademark of Advanced Sterilization Products a Johnson and Johnson Compan...

Page 3: ...mum filtration of fluids and contaminants and protecting the system from aspiration of these fluids In both systems the CO2 sensor generates infrared light and beams it through the airway adapter or s...

Page 4: ...es EtCO2 option Activate oxygen compensation if O2 levels in excess of 60 are introduced Activate N2O compensation if nitrous oxide is introduced into the airway circuit The presence of Desflurane bey...

Page 5: ...h windows in a vertical NOT a horizontal position This helps keep patient secretions from pooling on the windows Do NOT insert any object other than the sample cell into the sample cell receptacle on...

Page 6: ...t and property even if ZOLL has been advised of the possibility of such damages THE WARRANTY SET FORTH HEREIN IS EXCLUSIVE AND ZOLL MEDICAL CORPORATION EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES WHETHER...

Page 7: ...EtCO2 setup as follows Attaching the CAPNOSTAT 3 sensor cable Selecting a mainstream airway adapter Attaching the airway adapter to the CAPNOSTAT 3 sensor Zeroing the mainstream sensor airway adapter...

Page 8: ...ximately 15 seconds The unit displays the message ZERO DONE upon completion of the zeroing Note Do not attempt zeroing for 20 seconds after removing the adapter from the patient s airway This time all...

Page 9: ...ow steps a through c else go to step 4 a Verify proper connection of the adapter to the sensor b Verify that the airway adapter windows are clean and dry c If the adapter is properly connected and the...

Page 10: ...after its use When the previously used module is reconnected to the M Series unit module zeroing does not have to be repeated as long as no other LoFlo module has been connected and zeroed during the...

Page 11: ...lo module and ensure that it clicks into place 2 If the unit displays either of the following messages take the appropriate action 3 Place the airway adapter assembly at the proximal end of the airway...

Page 12: ...o module and rubber protective covering can be cleaned and disinfected by wiping with one or more of the following solutions 70 isopropyl alcohol 2 gluteraldehyde 10 bleach solution Use a soft cloth d...

Page 13: ...ftkey from the physiological monitoring menu cycles the display from the capnogram waveform to the plethysmograph waveform if SpO2 is installed to no second waveform displayed Use the Zoom softkey fro...

Page 14: ...airway circuit If the concentration of oxygen in the breathing circuit exceeds 60 and nitrous oxide is in use both O2 and N2O should be activated When the Comp softkey is pressed the unit displays the...

Page 15: ...alarm functions are disabled Background ECG analysis functions continue to operate as described in the AED section of the M Series Operator s Guide The ALARM SUSPEND button can be used to activate dea...

Page 16: ...ature unit displays WARM UP message 4 Perform a zero procedure if necessary see Zeroing the LoFlo CO2 Module on page 8 5 Breath normally into the adapter 6 Verify that the unit displays EtCO2 readings...

Page 17: ...3 kPa 0 95 mmHg OFF 0 12 5 OFF 0 12 6 kPa OFF High Respiration Rate Alarm Limit 120 respirations per min 5 150 respirations per min OFF Low Respiration Rate Alarm Limit 5 respirations per min 0 100 re...

Page 18: ...PU Oral Nasal CO2 sampling only pediatric Nasal CO2 Sampling with O2 Delivery Cannula Adult 8000 0356 SPU Nasal CO2 sampling with O2 delivery adult Nasal CO2 Sampling with O2 Delivery Cannula Pediatri...

Page 19: ...problem persists return for service CO2 DEVICE NOT READY The zero operation cannot be initiated because The sensor or module is still warming up No sensor or module is attached to the unit Wait for se...

Page 20: ...g the adapter type Clean airway adapter and reattach it If the problem persists or the adapter type was changed perform a mainstream airway adapter zero as described in Zeroing the Mainstream Sensor A...

Page 21: ...line is set below 0 mmHg Perform a sidestream module zero as described in Zeroing the LoFlo CO2 Module on page 8 CHECK CO2 LINE Sample line blockage Verify that the sample line is plugged into the mod...

Page 22: ...han 5 minutes Replace module ZERO CO2 MODULE Module cable connected to M Series unit for the first time Probe drift error detected or negative CO2 detected Perform module zero as described in Zeroing...

Page 23: ...lt pediatric adapters 50 ms Neonatal adapters 200 ms Respiration Rate Range 0 150 breaths per minute Respiration Rate Accuracy 1 breath per minute Sample Flow Rate N A 50 ml min Compensations Barometr...

Page 24: ...at maximum energy 360J or 1 5 hours minimum of continuous ECG monitoring or 1 0 hour of continuous ECG monitoring pacing at 60 mA 70 beats min For a new fully charged XL battery pack at 20 C 60 defibr...

Reviews: