Chapter 6. Battery Charger
32
www.zoll.com
Mode
Battery charger
indicator
Definition
Action
Charging
The battery charge
status indicator
flashes at a fast
rate.
The battery is charging. Typical
charge time is up to 2 hours.
Leave the battery in the battery char
-
ger. The battery charge indicator
shows the charge level. See “Battery
status” on page 24.
Measure
-
ment
cycle
The battery charge
status indicator
flashes at a slow
rate.
The battery is undergoing a
measurement cycle.
Leave the battery in the battery char
-
ger until the
measurement cycle com
-
pletes. See “Battery charger
measurement cycle” on page 32.
Out of
tempera
-
ture
range
The battery is either above or
below the temperature range.
Leave the battery in the battery char
-
ger until the
temperature returns to
range
. See “Using the battery char
-
Ready
The battery charge
status indicator
remains illuminated.
The battery is fully charged and
is ready to use.
Do one of the following:
• Leave the battery in the battery
charger to ensure a full charge
when needed.
• Install the battery in the plat
-
form.
• Store the battery in a cool dry
place.
Fail
The battery alert
indicator illumi
-
nates.
The battery charger was unable
to charge the battery or the bat
-
tery has failed the measurement
cycle.
Remove and reinsert the battery into
the battery charger. If the battery alert
remains illuminated, note the battery
serial number and contact "Technical
Support" (page
Idle
Neither the battery
charge status indi
-
cator nor the bat
-
tery alert indicator
illuminate.
The battery charger is unable to
recognize the battery.
Remove and reinsert the battery. If the
status is still Idle, see "Troubleshoot
-
ing the battery charger" on page
Table 6.1: Battery charger status indicators
Battery charger measurement cycle
The measurement cycle measures a battery’s capacity. If a battery has not undergone a full charge/discharge
cycle and is placed in the battery charger, a measurement cycle is automatically initiated.
The measurement cycle can take
5 to 10 hours
. Do not remove the battery from the battery charger until the
measurement cycle is complete, or the battery’s charge may be reduced. If the battery is removed before
the measurement cycle is complete, the next time the battery is inserted into the battery charger, the
measurement cycle automatically restarts.
After the measurement cycle completes, the battery is either ready to use (battery charge status indicator on
the battery charger panel illuminates) or has failed and should be replaced (battery alert indicator on the
battery charger panel illuminates).
If a battery has failed the measurement cycle, do not use it. Contact "Technical Support" (page
Summary of Contents for AutoPulse NXT
Page 1: ...14585 001 Rev 1 User Guide...
Page 9: ...Chapter 1 Introduction 4 www zoll com This page intentionally left blank...
Page 19: ...Chapter 3 Platform 14 www zoll com This page intentionally left blank...
Page 27: ...Chapter 4 Band 22 www zoll com This page intentionally left blank...
Page 31: ...Chapter 5 Battery 26 www zoll com This page intentionally left blank...
Page 39: ...Chapter 6 Battery Charger 34 www zoll com This page intentionally left blank...
Page 63: ...Chapter 9 Maintenance and Transport 58 www zoll com This page intentionally left blank...
Page 67: ...Appendix A System Checklist 62 www zoll com This page intentionally left blank...