• To satisfy FCC RF exposure requirements for mobile transmitting
devices, a separation distance of 20 cm or more should be
maintained between the antenna of this device and persons during
device operations. To ensure compliance, operations at closer than
this distance are not recommended. The antenna (located
internally in top front of device) used for this transmitter must not
be co-located in conjunction with any other antenna or transmitter.
Cautions
• The ZOLL AED 3 defibrillator may not perform to specifications
when stored at the upper or lower extreme limits of storage
temperature and then immediately put into use.
• Use only the ZOLL AED 3 nonrechargeable, disposable battery pack
with the ZOLL AED 3 defibrillator.
• Keep a new spare battery pack with the device at all times.
• When the unit displays the
CHANGE BATTERY
message, replace the
battery pack with a new one as soon as possible.
• Dispose of battery packs in accordance with federal, state, and local
regulations. Battery packs should be shipped to a reclamation
facility for recovery of metal and plastic compounds as the proper
method of waste management.
• ZOLL defibrillation pads contain no hazardous materials and may be
disposed of in general trash unless contaminated with pathogens.
Use appropriate precautions when disposing of contaminated
defibrillation pads.
• If the ZOLL AED 3 defibrillator is stored outside the recommended
environmental conditions, the defibrillation pads and/or battery
pack may be damaged or their useful life reduced.