Installation and Operation
9
Zolix Instruments Co., Ltd.
http://www.zolix.com.cn E-mail: [email protected]
5.2.2 CCD
CCD single spectrum range
Model
Grating
(
g/mm
)
Reciprocal
Dispersion
(
nm/mm
)
Mechanical
scanning
range (nm)
Resolution
(nm)-CCD
(
13.5um
)
Resolution
(nm)-PMT
Single
spectrum
range(nm)@
30mm CCD
Omni-λ200i
3600
1
0 to 400
0.1
0.05
30
2400
1.4
0 to 600
0.14
0.08
42
1800
2.16
0 to 800
0.2
0.1
64
1200
3.6
0 to 1200
0.28
0.15
108
900
4.96
0 to 1600
0.4
0.2
148
600
7.68
0 to 2400
0.56
0.3
230
300
15.77
0 to 4800
1.12
0.6
473
150
31.89
0 to 9600
2.24
1.2
956
★
The reference wavelength of the 3600 grating gratings in the table is 253.65nm, and the reference
wavelength of the rest of the grating gratings is 435.833nm, and the dispersion corresponding to
different wavelengths is different;
★
The spectral range is not only related to the mechanical scanning range, but also to the raster
shine wavelength
;
★
The resolution values listed in the table are not all given at the same wavelength, and the specified
detection wavelength is the standard wavelength near the wavelength of the blaze.
5.2.3 Shutter Operation
(
Options
)
1. The shutter in the spectrograph can be set by ZolixScan software, click on the
Monochrometer A
Digital Port.