N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
5-5
Pulse Oximetry Monitoring
5.5. Pulse Oximetry Monitoring Messages
Note
– Factory default SpO
2
alarm and limit settings may be found in the PPM3 Monitor
Settings chapter.
Message
Parameter
Value
Possible Cause
Suggested Action
SpO
2
< [lower limit]
[number]
The patient's oxygen
saturation has fallen
below the current lower
alarm limit.
Check the patient and provide any
necessary clinical care
Change the alarm limit if it is no longer
clinically appropriate
SpO
2
> [upper limit]
[number]
The patient's oxygen
saturation has risen
above the current upper
alarm limit.
Check the patient and provide any
necessary clinical care
Change the alarm limit if it is no longer
clinically appropriate
HR < [lower limit]
Note
– When PR is
sourced from SpO
2
the
Pulse Rate is labeled in
cyan as "HR (SpO2)" in
the HR parameter box.
PR alarm conditions
annunciate as HR
alarms.
[number]
The patient's pulse rate
has fallen below the
current lower alarm limit.
Check the patient and provide any
necessary clinical care
Change the alarm limit if it is no longer
clinically appropriate
HR > [upper limit]
Note
– When PR is
sourced from SpO
2
the
Pulse Rate is labeled in
cyan as "HR (SpO2)" in
the HR parameter box.
PR alarms use the HR
alarm limit settings. PR
alarm conditions
annunciate as HR
alarms.
[number]
The patient's pulse rate
has risen above the
current upper alarm limit.
Check the patient and provide any
necessary clinical care
Change the alarm limit if it is no longer
clinically appropriate
Summary of Contents for PPM3
Page 1: ...Nightingale Monitoring System PPM3 User s Guide ...
Page 4: ......
Page 16: ......
Page 28: ......
Page 32: ...2 4 Nightingale PPM3 User s Guide April 2019 Getting Started Figure 10 PPM3 Startup Screen ...
Page 52: ......
Page 96: ......
Page 100: ......
Page 138: ......
Page 160: ......
Page 164: ......
Page 184: ......
Page 194: ......
Page 196: ......
Page 198: ......