ZMD TSic LABkit User Manual And Technical Description Download Page 11

TSic LABkit™ User Manual and Technical Description

 

TSic

TM

 LABkit, Rev.2.4, Oct. 18, 2005 

Page 11 of 18 

© ZMD AG & IST AG, 2005 

All rights reserved. The material contained herein may not be reproduced, adapted, merged, translated, stored, or used without the prior 
written consent of the copyright owner.

 

6  Sensors for the TSic LABkit

TM

, Sensor Connection Cable and Connector 

6.1 

Standard Cables 

• 

Adaptor with TSic™ 306 E-Line Sensor Type 

• 

Length: 3 ft / 1m 

• 

Maximum temperature: 80°C  

• 

Blocking capacitor: 100nF connected close to the sensor between VDD and GND 

 

Note:

 If the standard cable requires an extention, add it on the connector side of the cable to minimize the 

distance between the blocking capacitor and the sensor. 
 

Order Number 
for sensor with cable 

TEFÜ-TSIC306-IST 
 

Sensor 

TSic™ 306 

Temperature Range 
with delivered cabeling 

-20°C to 80°C 
(Short-term temperature  
strain: -50°C to 120°C) 

Protection 

Sensor enameled 

Connection Cable 

PVC connection wire, 3ft (1m) 
4-pole, unshielded, t

max

 80°C 

Connector 

3-pole, nylon connector, 
soldered, crimped, connectors 
not shrunken 

Pin 

V+ (VDD)  yellow 
DATA 

green 

GND 

white 

 

 

Summary of Contents for TSic LABkit

Page 1: ...WARE 10 5 1 USB HARDWARE CONFIGURATION 10 5 2 MICROCONTROLLERS IN CIRCUIT PROGRAMMING 10 6 SENSORS FOR THE TSIC LABKIT TM SENSOR CONNECTION CABLE AND CONNECTOR 11 6 1 STANDARD CABLES 11 6 2 REQUIREMEN...

Page 2: ...wner 1 TSic LABkit Contents a CD ROM with the TSic Recorder Software for Microsoft Windows b TSic LABkit TM Adapter for USB c Four TSic TM 306 Sensors with Cable Connector d USB Cable Figure 1 1 LABki...

Page 3: ...rface USB Adapter Dimensions 80 mm x 40 mm x 23 mm EMC Noise Emission Noise Immunity 89 336 EEC European Harmonized Standards EN 61000 6 3 2001 EN 61000 6 2 2001 Note The technical data provided above...

Page 4: ...soft Excel 4 1 System Requirements and Software Installation System Requirements 5x86 compatible PC 32 MB RAM Hard drive with 20MB free space USB port Microsoft W98 ME 2000 XP Internet Explorer 6 0 re...

Page 5: ...after installation is for the TSic 306 If other TSic types are connected see section 4 4 2 below for required sensor type configuration using the Settings dialog window The Settings dialog window can...

Page 6: ...urements is selected in the Settings dialog window See section 4 4 The Raw Value readings are the decimal values output by the TSic sensors Conversion into C or F values is based on the conversion tab...

Page 7: ...in the main window To close the Settings dialog window and apply any new settings click OK Figure 4 2 TSic Recorder Settings Dialog Window Figure 4 3 USB COM Port Setting 4 4 1 USB COM Port Settings I...

Page 8: ...figure the correct measurement range Note In addition to the standard sensor products offered by IST customized OEM sensor types are also available from IST and can be manually configured as specified...

Page 9: ...ta is recorded enter the new directory and filename which must be write accessible in the Path dialog box Default directory and filename C TSIC LABKIT DATA tempdata csv Change Filename at 12 00 a m Se...

Page 10: ...protection 5 1 USB Hardware Configuration The jumpers on the PCB control the following settings Important Connect either JP1 or JP8 not both at the same time Pin Connections Name Settings JP1 1 2 GND...

Page 11: ...imum temperature 80 C Blocking capacitor 100nF connected close to the sensor between VDD and GND Note If the standard cable requires an extention add it on the connector side of the cable to minimize...

Page 12: ...sor between VDD and GND 3 pole connector Schukat Electronics part no NSG25 3 crimp contact NSK25 0 7 Electromagnetic Compatibility EMC Standard 89 336 EU The TSic LABKIT USB Adapter meets the electrom...

Page 13: ...sent continuously and separated by semicolon The character and carriage return CR mark the end of the transmission Typical data line structure example 08DA 08C2 08CC 08CA 2332 CR Characters Definition...

Page 14: ...0 C 0x7FF 07FF XXXX XXXX XXXX CCCC CR Not valid 0x800 to 0xFFF UUUU XXXX XXXX XXXX CCCC CR No response Timeout no sensor FFFF XXXX XXXX XXXX CCCC CR XXXX Other channels CCCC Value for check sum UUUU V...

Page 15: ...C accuracy 0 1 C for 5 C to 45 C Test Messung 1 Text2 Text3 Sensoren 4x TSic 306 Lauf Nr Date Time Sensor 1 Sensor 2 Sensor 3 Sensor 4 2897 08 08 2005 17 39 39 24 65 24 74 24 74 24 55 2898 08 08 2005...

Page 16: ...23 02 22 23 58 22 24 54 22 25 51 22 26 47 22 27 43 22 28 39 T1 T2 T3 T4 57 57 5 58 58 5 59 59 5 60 60 5 61 61 5 62 62 5 63 63 5 22 14 42 22 14 52 22 15 02 22 15 12 22 15 22 T1 T2 T3 T4 Example 1 Proc...

Page 17: ...in the accuracy range The chart shows the higher resolution of TSic 506F which is shown as sensor T4 Figure 8 6 Surface Temperature Measurement Before and After Touching the Sensors Number Date Time...

Page 18: ...licensee or third party for any damages including but not limited to personal injury property damage loss of profits loss of use interruption of business or indirect special incidental or consequentia...