1. SAFETY

Please read and understand all safety hazards and warnings prior to using your Hot Spot 110 

furnace. If you have any questions about this section please contact Zircar Zirconia, Inc. 

BEFORE

 

start up.

ELECTRICAL WARNINGS

1.

  The Hot Spot 110 is designed to be powered by a single phase circuit breaker or fuse protected 

grounded outlet and is supplied with a cord with a plug appriopriate for the design voltage.

  Operating with an improper power supply could cause the cord to overheat and cause a fire.

  No other device should be operated on the circuit used by the furnace. The furnace works 

best over the long run if supplied with a constant voltage.

  Please refer to the furnace nameplate to verify your Hot Spot power requiprements and 

reference the chart below.

Model

Voltage

Amperage

Hot Spot 110

110-120

20

Hot Spot 110

220-240

10

Hot Spot 110L

110-120

20

Hot Spot 110L

220-240

10

2.

  Disconnect all power from the furnace before opening any part of the Hot Spot 110 cabinet. 

Failure to disconnect the power supply can result in electrocution. Unplug the power cord to 

disconnect the power.

3.

  The Hot Spot 110 is electrically resistance heated by exposed heating elements. NEVER TAMPER 

with the safety cut off switch. The safety cutoff switch cuts off power to the heating elemenets 

when the door is opened (see appendix A #8).

BURN/FIRE WARNINGS

1.

  Do not locate the Hot Spot 110 near combustible materials. Select a safe location to install the 

furnace.

  The furnace door opens up and away from the operator and can be opened when very hot. If 

opened while hot, the door will radiate heat over the top of the furnace. Combustibles 

located in the path of this radiated heat may catch fire.

1

Summary of Contents for Hot Spot 110

Page 1: ...Hot Spot 110 User Guide Zircar Zirconia Inc www zircarzirconia com R 110M 01 Rev00...

Page 2: ...Temperature Controller Configuration D REX P48 Temperature Controller Patterns Programs E CL100L Over Temperature Controller Configuration F SCR G REX P48 96 Series Programmable Temperature Controlle...

Page 3: ...eference the chart below Model Voltage Amperage Hot Spot 110 110 120 20 Hot Spot 110 220 240 10 Hot Spot 110L 110 120 20 Hot Spot 110L 220 240 10 2 Disconnect all power from the furnace before opening...

Page 4: ...gh temperature for a prolonged time and then cooled down may have internal stresses which can cause the protective silica glaze to splinter into small fragments There have been instances where heating...

Page 5: ...e furnace chamber Typical contaminates are Previously processed material remnants Glaze which has fallen from the heating elements 5 An appropriate firing setter should always be used to protect the H...

Page 6: ...ning the handle and lifting outwards and upwards until it catches on the latch see appendix A 7 Remove the packing material from the heating chamber Be careful not to touch the heating elements IMPORT...

Page 7: ...ed position 5 Switch the green power switch to the ON position see appendix A 3 The light will turn green the fan will come on and the controllers will start up 6 Push the PTN1 PTN2 button on the prog...

Page 8: ...revised If you MUST reduce the OT limit it can be reduced as follows 1 Press the SET button The rightmost digit will be highlighted in the lower orange display see appendix A 36 2 Press the v or butt...

Page 9: ...er step 8 above will cause the furnace to power off at the end of that step Step 1 of the pattern will start at the actual temperature of the furnace when the pattern is run Example If the end point a...

Page 10: ...bration and should be replaced For video instructions on how to replace the thermocouple go to www zircarzirconia com technical documents procedure replacing hotspot thermocouple Heating Elements Even...

Page 11: ...tion 3 The furnace does not heat and the HEAT ON indicator see appendix A 5 does not illuminate Check that the Heat switch is in the on position which is position l see appenfix A 4 Check that the Ove...

Page 12: ...nd braid connections were all good the meter should show low ohms which indi cates a dead short 5 The over temperature controller alarms during a firing Check the over temperature controller set point...

Page 13: ...NY 10921 0827 www zircarzirconia com Phone 845 651 3040 Fax 845 651 0074 email sales zircarzirconia com Technical Information Sales Manager David C Hoskins Phone 845 651 3040 ext 137 Fax 845 651 0074...

Reviews: