background image

Hot Spot 110

User Guide

Zircar Zirconia, Inc.  

 

www.zircarzirconia.com

R-110M-01 Rev00

Summary of Contents for Hot Spot 110

Page 1: ...Hot Spot 110 User Guide Zircar Zirconia Inc www zircarzirconia com R 110M 01 Rev00...

Page 2: ...Temperature Controller Configuration D REX P48 Temperature Controller Patterns Programs E CL100L Over Temperature Controller Configuration F SCR G REX P48 96 Series Programmable Temperature Controlle...

Page 3: ...eference the chart below Model Voltage Amperage Hot Spot 110 110 120 20 Hot Spot 110 220 240 10 Hot Spot 110L 110 120 20 Hot Spot 110L 220 240 10 2 Disconnect all power from the furnace before opening...

Page 4: ...gh temperature for a prolonged time and then cooled down may have internal stresses which can cause the protective silica glaze to splinter into small fragments There have been instances where heating...

Page 5: ...e furnace chamber Typical contaminates are Previously processed material remnants Glaze which has fallen from the heating elements 5 An appropriate firing setter should always be used to protect the H...

Page 6: ...ning the handle and lifting outwards and upwards until it catches on the latch see appendix A 7 Remove the packing material from the heating chamber Be careful not to touch the heating elements IMPORT...

Page 7: ...ed position 5 Switch the green power switch to the ON position see appendix A 3 The light will turn green the fan will come on and the controllers will start up 6 Push the PTN1 PTN2 button on the prog...

Page 8: ...revised If you MUST reduce the OT limit it can be reduced as follows 1 Press the SET button The rightmost digit will be highlighted in the lower orange display see appendix A 36 2 Press the v or butt...

Page 9: ...er step 8 above will cause the furnace to power off at the end of that step Step 1 of the pattern will start at the actual temperature of the furnace when the pattern is run Example If the end point a...

Page 10: ...bration and should be replaced For video instructions on how to replace the thermocouple go to www zircarzirconia com technical documents procedure replacing hotspot thermocouple Heating Elements Even...

Page 11: ...tion 3 The furnace does not heat and the HEAT ON indicator see appendix A 5 does not illuminate Check that the Heat switch is in the on position which is position l see appenfix A 4 Check that the Ove...

Page 12: ...nd braid connections were all good the meter should show low ohms which indi cates a dead short 5 The over temperature controller alarms during a firing Check the over temperature controller set point...

Page 13: ...NY 10921 0827 www zircarzirconia com Phone 845 651 3040 Fax 845 651 0074 email sales zircarzirconia com Technical Information Sales Manager David C Hoskins Phone 845 651 3040 ext 137 Fax 845 651 0074...

Reviews: