Page 29
CLEANING & MAINTENANCE
ZiXpress 32 Dx
system Cleaning
◼
After every operation of purification, please properly
remove used plastic consumables.
◼
At the end of every working day, please wipe clean any
fluid that remains on the inside and outside of the
instrument with damp cloth, use 75% ethanol and UV
light to disinfect the surface.
Note:
Press
button to get into the UV light mode, and
select the disinfecting time.
◼
Clean the instrument body by removing dust gently with a
dry, soft cloth. If the outside of the
ZiXpress 32 Dx
system
is heavily soiled, or if any samples that may
cause infection are adhered to the outside of the
ZiXpress 32 Dx system
, wipe with a soft paper tissue,
soaked with 75% ethanol.
ZiXpress 32 Dx
system
Maintenance
Two types of maintenance have to be performed on the
ZiXpress 32 Dx system
instrument, for details on each
type of maintenance, see the list below:
Maintenance Type
Performed by
Schedule
Routine
Cleaning instrument
body (outside)
Cleaning instrument
body (inside)
User
User
Bi-weekly
Daily
Preventive
Service Engineer Annually