www.zinexts.com 27 MagPurix 24 EVO Instruction Manual
(1)
UV Light 30 mins
Clean the Sample Tray and M motor with mild detergent and rinse with deionized
water. Allow the parts to dry before use. Clean and disinfect the platform surface
by wiping with deionized water followed by 75% ethanol.
Clean the instrument body by removing dust gently with a dry, soft cloth. If the
outside of the MagPurix 24 EVO system is heavily soiled, or if any sample that
may cause infection are adhering to the outside of the MagPurix 24 EVO system,
wipe with a soft paper tissue, soaked with 75% ethanol.