![ZINEXTS MagPurix 24 EVO Instruction Manual Download Page 26](http://html.mh-extra.com/html/zinexts/magpurix-24-evo/magpurix-24-evo_instruction-manual_934159026.webp)
www.zinexts.com 26 MagPurix 24 EVO Instruction Manual
3.2 Routine and Preventive Maintenance
Two types of maintenance are to be performed on the MagPurix 24 EVO system
instrument as listed in the table. For details on each type of maintenance, see below:
Schedule
Maintenance Type
Performed by
After each use
Routine
Cleaning of sample Tray, M
module, piston and holder lid
by 75% EtOH.
Cleaning of heating module
and using cotton swab to check
buffer residue in hold of heating
module.
User
Bi-weekly
Check SD Memory space
Cleaning of instrument body
(outside)
Cleaning of instrument body
(inside)
User
Two times a year or
Annually
Preventive
Grease ball screws if required
Check motor movement
Check the air-tight status with
piston units (using empty plastic
wares and run )
Cleaning of Piston
UV light change annually
Service Engineer