94
Paralysis diagnostics
19.3 Chronaxia/accommodation
quotient
19
Step 3: Accept rheobase value.
By accepting the calculated pulse, the intensity is automatically set to double
the rheobase value.
The intensity cannot be modified in this step.
Chronaxie
Pulse duration for triggering a minimum muscle twitch at a surge current
strength that is double the rheobase value.
Calculating
chronaxia
Step 1: Increase pulse time in stages (or reduce if required)
Step 2: Trigger a single pulse.
Repeat steps 1 and 2 until a minimum muscle twitch is observed.
The chronaxie value is calculated using the first minimum muscle twitch.