Electromagnetic compatibility (EMC)
table
8
Page 35
This device is suitable for use in a “medical environment”.
This device meets the requirements of IEC 60601-1-2.
Precautionary measures, compliance with the EMC guideline information contained in
this manual, and the review of all medical devices in simultaneous operation are
necessary in order to ensure the electromagnetic compatibility and coexistence of all
other medical devices in the immediate vicinity prior to a treatment procedure.
WARNING
The use of this device next to or stacked with other devices should be avoided since
this could lead to malfunctioning of the device. If such use is necessary, this device
and the other devices should be observed to ensure that they are working normally.
WARNING
The use of accessories and cables other than those specified or provided by the
manufacturer of this device can lead to increased electromagnetic emissions or
decreased electromagnetic immunity of this device and to faulty operation.
WARNING
Portable HF communication devices (including peripheral devices such as antenna
cables and external antennas) should be used at a minimum distance of 30 cm (12
inches) from any part of the device, including the cable specified by Eunsung Global
Corp. Otherwise the performance may be impaired.
WARNING
The device is suitable for use in all establishments except for private households and
in those which are connected directly to the public low-voltage grid.
Note
The EMISSIONS properties of this device make it suitable for use in industrial areas
and hospitals (CISPR 11 class A).
When it is used in a home environment (for which CISPR 11 class B is normally
necessary), this device may not offer adequate protection for high frequency. The user
may need to take additional protective measures for safety, such as moving or
reconnecting the device.