Operation instruction
9.1 Performing treatments
9
Page 15
Activate the laser
Pressing the "Start" button makes the laser ready to operate.
This is indicated with the message "Ready" in the footer. If the "Start" button is
no longer lit up (because it has already been pressed), the "Stop" button
becomes active.
Start treatment
By pressing the foot switch, the laser beam is delivered and the "Ready"
message switches to "Active".
Note:
During use, the patient must be monitored carefully and the treatment should
be adjusted or terminated as necessary if any issues arise.
End treatment
De-activating the foot switch cuts off the laser beam.
The message in the footer switches from "Active" back to "Ready".
The output must be manually reset to 0.0 W. Use the controller to set the
output level.
Note:
If output is terminated with the "Stop" button, both the "Start" button and the
foot switch must be activated to restart treatment.
Place the applicator back into the calibration sensor once treatment is
complete.