Operation instructions
7.2 Performing the treatment
7
Page 10
Note
:
The following descriptions are based on the factory settings.
Note:
All buttons, menus and sub-menus can be activated directly on the screen by
pressing with the finger.
Open therapy screen
Activating the “Therapy” button opens the therapy screen.
Select gel pad
Select the appropriate gel pad and place this correctly into the handpiece.
Note:
To guarantee optimal energy transfer, use sufficient conductive gel between
the handpiece and the gel pad.
Adjust parameters
Adjust the desired energy level, number of shocks and frequency by selecting
the parameter in question and adjusting the value using the central control
knob.
If the optional foot switch is connected, the “-” and “+” pedals can be used for
adjustment as well.
Apply conductive
gel
Apply the conductive gel to the treatment area and spread the gel evenly over
the entire surface of the treatment area
Start of therapy
Note:
Note:
Start a treatment by selecting the “Start” button, immediately followed by
clicking the orange button on the handpiece (19) or pressing the “Shock” pedal
on the optional foot switch.
If the "Start" button is not activated, no power will be emitted
.
The energy level and frequency can be increased or decreased during therapy
by selecting the parameter in question followed by using the central control
knob or the optional foot switch.
End of therapy:
Note:
An acoustic signal will sound after the therapy time has elapsed.
If the volume is set to “0” in the “Settings” menu, no acoustic signal will sound.