Functional Test
15
Page 25
Routine inspection of
equipment
The covering of power line of equipment, transducer connecting line, etc. shall not
be peeled off and internal lines shall not be exposed, and shall not be damaged by
impact from outside.
There shall be no trace of oil leakage from transducer.
Wash the outside of equipment so that there is no foreign material.
The button for equipment operation, etc. must not shake.
The various parts attached to the device must not shake.
If any of the above occurs, contact your servicepartner for help.
Safety inspection
In order to ensure safe use, internal cleaning should be performed once per year by
a person authorized by Zimmer MedizinSysteme GmbH
In order
to ensure safe use, be sure to check the equipment including internal
components and output voltage from the person who has been given authorization
from the company once per year.
Please clean the transducer before storing
After storing the device for a long period of time, be sure to check the device before
using it.
Please note the following regarding storage conditions:
Keep out of water
Keep away from direct sunlight
Do not store near heaters
Avoid locations subject to excessive shock or vibration, exposure to
chemicals or explosive gases.
Troubleshooting
If the equipment does not operate normally during use, please check the items listed
in the table before requesting service. If none of the following problems apply, or if
the following remedies do not help, turn off the power to the equipment and contact
your servicepartner.