DCO-1318 

 

Effective Date: 01-SEP-2021

 

LBL 6028 Rev. G 

Page 9 of 23 

 

10. 

SYSTEM SETUP 

CONTROLLER SET UP INSTRUCTIONS 

1.

 

Place the Tricera Controller on a stable flat work surface.   

 

IMPORTANT: Prior to use verify that the Controller and footswitch 

are decontaminated and clean and that the shaver handpiece is 

clean and sterilized.  

2.

 

Connect the Tricera Controller power cord (Fig.9) to the power entry 

module (Fig. 2, Item 10).    

 

 

Figure 9

 

a.

 

Optionally, set up the ZEOS system by connecting the ZEOS 

console to the rear of the Tricera Controller via the USB port 

(Fig. 2, item 12). See Appendix B for ZEOS interface details. 

3.

 

Connect the Tricera footswitch cable to the footswitch receptacle 

(Fig. 1, Item 1) on the left-hand side of the front panel of the 

Controller (Fig. 10). 

 

 

Figure 10

 

4.

 

Turn on the Tricera Controller using the power switch (Fig. 1, Item 

7) on the front of the Controller.    
a.

 

Optionally, open the profile screen and select a user profile 

(See Appendix H for user profile details). 

5.

 

The Zimmer Biomet logo and self-test progress will appear on the 

screen, followed by the images shown in Figure 11 and 12.   

 

Figure 11

 

  

 

Figure 12

 

6.  Open the FMA Cassette packaging and dump the contents onto the 

sterile table.   

 

When using Veriflow (R-10003):

 Within the sterile field, retain the 

patient ends of the Inflow and suction connectors (red and blue 

connectors). Pass the FMA cassette out of the sterile field with the 

saline spikes and waste outflow tube. 

 

When using Exoflow (R-10017):

 Within the sterile field, retain the 

patient end of the suction connector (red connector). Pass the FMA 

cassette out of the sterile field with the waste outflow tube. 
 

Connect the FMA cassette to the controller pumps by placing the 

cassette in place as shown.  The controller will detect the presence of 

the FMA cassette and will provide an audible tone when the cassette is 

in the correct position.  
 
 
 
 
 
 
 
 
 

Figure 13 

7.  Press the lock button and the FMA cassette will automatically lock 

into place.  

 

NOTE: for Exoflow Outflow FMA (R-10017) only, the 

Controller will proceed to the screen below after locking; no 

purge/calibration is necessary. 

Summary of Contents for Tricera

Page 1: ...15 Appendix C Abbreviations 15 Appendix D Tones 15 Appendix E Troubleshooting 16 Appendix F Essential Performance Power Curve 18 Appendix G EMC Tables 19 Appendix H User Profiles 21 1 SYMBOLS USED ON...

Page 2: ...on during diagnostic and operative arthroscopic procedures The Shaver Blade RF Probe of the Tricera Arthroscopic System provides abrasion resection debridement and removal of bone through its shaver b...

Page 3: ...hin the joint The inflow stopcock should be completely open or completely closed WARNING Keep the pump dry If liquid has accidentally leaked into the pump from the Fluid Management Accessories cassett...

Page 4: ...ble devices such as cardiac pacemakers If the patient has an implantable cardioverter defibrillator ICD contact the ICD manufacturer for instructions before performing ablation coagulation or other el...

Page 5: ...ay result in damage to the tip of the devices 7 ENVIRONMENTAL PROTECTION Observe all national waste management regulations DO NOT dispose of Waste Electrical and Electronic Equipment WEEE as unsorted...

Page 6: ...6 Receptacle reserved for future functionality 7 Power ON OFF Switch and Power ON LED 8 Pressure Sensors Figure 2 Controller Rear 9 Equipotential Lug 10 Power Entry Module 11 Fuse Drawer 12 USB Acces...

Page 7: ...nt 6 Inflow to Patient 7 Inflow Pressure Diaphragm 8 Inflow Peristaltic Pump Tube 1 TPE Pump Tubing 2 Barb Lock 3 PVC Waste Tuing 4 Outflow Lumen 5 Barb Connector 6 Cassette Magnet Figure 5 Veriflow T...

Page 8: ...Date 01 SEP 2021 LBL 6028 Rev G Page 8 of 23 Figure 7 Tricera Handpiece LCD and Buttons 1 Activate 2 Coagulation 3 Joystick 4 Mode 9 5 Shaver Blades 1 Coupler 2 Hub 3 Shaft 4 Tip Figure 8 Tricera Devi...

Page 9: ...x H for user profile details 5 The Zimmer Biomet logo and self test progress will appear on the screen followed by the images shown in Figure 11 and 12 Figure 11 Figure 12 6 Open the FMA Cassette pack...

Page 10: ...e end of the Tricera handpiece as shown Figure 18 NOTE steps 14 17 apply to Veriflow FMA set R 10003 only 14 Select the joint for the procedure by touching the joint selection touchscreen or using the...

Page 11: ...at any time during the procedure 2 The Tricera Controller will set the following default pressure set point based on the initial joint selection Knee 45mmHg Shoulder 50mmHg Hip 50mmHg Small Joint 35m...

Page 12: ...e the potential of blood and debris drying prior to cleaning 3 After use Tricera Devices may be a potential biohazard and should be handled in accordance with acceptable medical practice and applicabl...

Page 13: ...5 4 Allow 0 Suction YES and NO When this setting is set to Yes the suction level for Oscillate Bone Cutting Bone Burring Forward and Reverse all non RF modes shall be adjustable between 0 and 100 When...

Page 14: ...of merchantability and fitness for a particular purpose and all other obligations and liabilities on the part of Zimmer Biomet Zimmer Biomet neither assumes nor authorizes any other person to assume f...

Page 15: ...Pressure 878 to 1082 cmH2O 86 to 106 kPa IV Transport and Storage Requirements Temperature 18 C to 60 C Relative Humidity 15 to 85 non condensing Atmospheric Pressure 510 to 1082 cmH2O 50 to 106 kPa A...

Page 16: ...tomer service E11 E16 Inflow turns off High pressure detected on the inflow line Foreign object between diaphragm and pressure sensors Controller malfunction Remove any obstructions from inflow line o...

Page 17: ...t the footswitch is plugged into the gray port on the Controller Ensure the Handpiece is securely plugged into the connector Replace Tricera Device N A N A N A F12 Handpiece Cannot Position Window Pro...

Page 18: ...ntentional activation of RF output no unintentional activation of pump motors and correct pressure indication within one indicator bar II Power Curve Full and Half Output Power versus Load Resistance...

Page 19: ...ootswitch 10 ft Hospital Grade Power Cord Table 1 Electromagnetic Emissions Statement Guidance and Manufacturer s Declaration Electromagnetic Emissions The Tricera System is intended for use in the el...

Page 20: ...lity should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 4...

Page 21: ...3 79 3 79 7 27 100 12 0 12 0 23 0 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to...

Page 22: ...3000 6000 Forward RPM Reverse Default 1000 6000 6000 Reverse RPM Bone Cutting Default 3000 15000 15000 Bone Cutting RPM Bone Burring Default 3000 15000 15000 Bone Burring RPM RF Ablation Default 1 10...

Page 23: ...handpiece shall continue running when the button is released When a button is pressed again the motor shall stop running In RF modes or when Handpiece Button Latch non RF modes is Off activation shall...

Reviews: