Zimmer Biomet Tricera Instruction Manual Download Page 5

 

DCO-1318 

 

Effective Date: 01-SEP-2021

 

LBL 6028 Rev. G 

Page 5 of 23 

 

 

Use only normal saline (sodium chloride 0.9% w/v; 150mmol/l) or 

Ringer’s lactate irrigation solution.  The performance of the system 

can be adversely affected by use of any other solution. 

 

DO NOT pinch, step on, kink or otherwise occlude the tubing set.  

Tubing restrictions can result in high pressure or poor device 

performance. 

 

Continuous, extended RF energy output may cause the Tricera 

Controller to overheat.  If this occurs, the Controller must be 

allowed to cool down before further use. 

Shaver Blade/Burr and RF Probe Precautions 

 

Excessive force on the shaver blade/burr and RF probe tip does 

not improve the device’s performance and may increase the risk of 

tissue and/or device damage.

 

 

DO NOT allow the tip of the shaver blades/burrs and RF probes to 

touch any hard object.  If such contact does occur, inspect the tip.  

If there are cracks, fractures, or if there is any other reason to 

suspect the tip is damaged, replace the device immediately.

 

 

Any monitoring electrodes are recommended to be placed as far 

as possible from the Tricera devices when high frequency surgical 

equipment and physiological monitoring equipment are used 

simultaneously on the same patient.  Monitoring systems 

incorporating high frequency current-limiting devices are 

recommended for use.   

 

Excessive force applied during insertion or removal of the shaver 

blades/burrs and RF probes may result in device damage or tissue 

injury.

 

 

Insertion and removal of the shaver blades/burrs and RF probes 

should always be under direct visualization.

 

 

DO NOT activate the shaver blades/burrs and RF probes unless 

the tip is immersed in a saline environment.  Electrodes may arc if 

activated in air, damaging the device.

 

 

DO NOT use the shaver blades/burrs and RF probes as a tool for 

mechanical displacement of tissue or bone or use the devices as a 

prying or scrubbing tool, as this may result in damage to the tip of 

the devices. 

7. 

ENVIRONMENTAL PROTECTION

 

 

Observe all national waste management regulations. 

 

DO NOT dispose of Waste Electrical and Electronic Equipment 

(WEEE) as unsorted municipal waste. 

 

Follow local governing ordinances and hospital practice regarding 

the disposal of the Probe/Tricera Devices and Fluid Management 

Accessories – Disposable Devices. 

 

The Controller contains an electronic printed circuit assembly.   At 

the end of the useful life of the equipment, it should be disposed of 

in accordance with any applicable national or institutional policy 

relating to obsolete electronic equipment. 

8. 

HOW SUPPLIED 

8.1.

 

Controller with Integrated Fluid Management 

 

The Controller is supplied non-sterile and is intended for reuse as a 

capital equipment in an orthopaedic or arthroscopic surgical 

environment.

 

The shipping box contains: 

 

One (1) Controller  

 

One (1) Footswitch  

 

One (1) Detachable 10 ft. Hospital Grade Power Cord 

 

One (1) Tricera System Instructions Manual 

8.2.

 

Fluid Management Accessories (All Models)

 

The Fluid Management Accessories are supplied sterile and 

are intended for single-patient use only.  DO NOT re-sterilize 

or reuse.  After each patient use, dispose of properly. 
The shelf box contains: 

 

One (1) Fluid Management Accessories 

 

One (1) Tricera System Package Insert 

8.3.

 

Shaver Blade/Burr and RF Probe (All Models) 

The shaver blades/burrs and RF probes are supplied sterile 

and are intended for single patient use only.  DO NOT re-

sterilize or reuse.  After each patient use, dispose of properly. 
The shelf box contains: 

 

Four (4) Shaver Blade/Burr or RF Probe  

 

One (1) Tricera System Package Insert 

8.4.

 

Tricera Handpieces (All Models) 

The Tricera Handpiece is supplied non-sterile and is intended 

for reuse.  It must be cleaned and sterilized before the first 

use. Prior to each subsequent use, it must be cleaned and 

sterilized. 
The shipping box contains: 

 

One (1) Tricera Handpiece  

 

One (1) Tricera Handpiece Instructions for Use 

(including cleaning and sterilization instructions) 

 

Three (3) Cleaning Brushes 

o

 

One (1) Tricera Device Cavity Brush 

o

 

One (1) Motor Shaft Brush 

o

 

One (1) Fluid Channel Brush 

 
 
 
 
 
 

 

 

Summary of Contents for Tricera

Page 1: ...15 Appendix C Abbreviations 15 Appendix D Tones 15 Appendix E Troubleshooting 16 Appendix F Essential Performance Power Curve 18 Appendix G EMC Tables 19 Appendix H User Profiles 21 1 SYMBOLS USED ON...

Page 2: ...on during diagnostic and operative arthroscopic procedures The Shaver Blade RF Probe of the Tricera Arthroscopic System provides abrasion resection debridement and removal of bone through its shaver b...

Page 3: ...hin the joint The inflow stopcock should be completely open or completely closed WARNING Keep the pump dry If liquid has accidentally leaked into the pump from the Fluid Management Accessories cassett...

Page 4: ...ble devices such as cardiac pacemakers If the patient has an implantable cardioverter defibrillator ICD contact the ICD manufacturer for instructions before performing ablation coagulation or other el...

Page 5: ...ay result in damage to the tip of the devices 7 ENVIRONMENTAL PROTECTION Observe all national waste management regulations DO NOT dispose of Waste Electrical and Electronic Equipment WEEE as unsorted...

Page 6: ...6 Receptacle reserved for future functionality 7 Power ON OFF Switch and Power ON LED 8 Pressure Sensors Figure 2 Controller Rear 9 Equipotential Lug 10 Power Entry Module 11 Fuse Drawer 12 USB Acces...

Page 7: ...nt 6 Inflow to Patient 7 Inflow Pressure Diaphragm 8 Inflow Peristaltic Pump Tube 1 TPE Pump Tubing 2 Barb Lock 3 PVC Waste Tuing 4 Outflow Lumen 5 Barb Connector 6 Cassette Magnet Figure 5 Veriflow T...

Page 8: ...Date 01 SEP 2021 LBL 6028 Rev G Page 8 of 23 Figure 7 Tricera Handpiece LCD and Buttons 1 Activate 2 Coagulation 3 Joystick 4 Mode 9 5 Shaver Blades 1 Coupler 2 Hub 3 Shaft 4 Tip Figure 8 Tricera Devi...

Page 9: ...x H for user profile details 5 The Zimmer Biomet logo and self test progress will appear on the screen followed by the images shown in Figure 11 and 12 Figure 11 Figure 12 6 Open the FMA Cassette pack...

Page 10: ...e end of the Tricera handpiece as shown Figure 18 NOTE steps 14 17 apply to Veriflow FMA set R 10003 only 14 Select the joint for the procedure by touching the joint selection touchscreen or using the...

Page 11: ...at any time during the procedure 2 The Tricera Controller will set the following default pressure set point based on the initial joint selection Knee 45mmHg Shoulder 50mmHg Hip 50mmHg Small Joint 35m...

Page 12: ...e the potential of blood and debris drying prior to cleaning 3 After use Tricera Devices may be a potential biohazard and should be handled in accordance with acceptable medical practice and applicabl...

Page 13: ...5 4 Allow 0 Suction YES and NO When this setting is set to Yes the suction level for Oscillate Bone Cutting Bone Burring Forward and Reverse all non RF modes shall be adjustable between 0 and 100 When...

Page 14: ...of merchantability and fitness for a particular purpose and all other obligations and liabilities on the part of Zimmer Biomet Zimmer Biomet neither assumes nor authorizes any other person to assume f...

Page 15: ...Pressure 878 to 1082 cmH2O 86 to 106 kPa IV Transport and Storage Requirements Temperature 18 C to 60 C Relative Humidity 15 to 85 non condensing Atmospheric Pressure 510 to 1082 cmH2O 50 to 106 kPa A...

Page 16: ...tomer service E11 E16 Inflow turns off High pressure detected on the inflow line Foreign object between diaphragm and pressure sensors Controller malfunction Remove any obstructions from inflow line o...

Page 17: ...t the footswitch is plugged into the gray port on the Controller Ensure the Handpiece is securely plugged into the connector Replace Tricera Device N A N A N A F12 Handpiece Cannot Position Window Pro...

Page 18: ...ntentional activation of RF output no unintentional activation of pump motors and correct pressure indication within one indicator bar II Power Curve Full and Half Output Power versus Load Resistance...

Page 19: ...ootswitch 10 ft Hospital Grade Power Cord Table 1 Electromagnetic Emissions Statement Guidance and Manufacturer s Declaration Electromagnetic Emissions The Tricera System is intended for use in the el...

Page 20: ...lity should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 4...

Page 21: ...3 79 3 79 7 27 100 12 0 12 0 23 0 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to...

Page 22: ...3000 6000 Forward RPM Reverse Default 1000 6000 6000 Reverse RPM Bone Cutting Default 3000 15000 15000 Bone Cutting RPM Bone Burring Default 3000 15000 15000 Bone Burring RPM RF Ablation Default 1 10...

Page 23: ...handpiece shall continue running when the button is released When a button is pressed again the motor shall stop running In RF modes or when Handpiece Button Latch non RF modes is Off activation shall...

Reviews: