TR250
12410-0701-03
Page
4 / 18
www.ziehl.de
6
button Set / Reset ►
Press short
Display next sensor
Press short in menu mode
switch to programming mode, the right decimal-point flashes
Press short in
programming mode
switch to next point, the set value will be stored
Press for 10 s when switching on supply voltage until displayed
„
Pr1
“
. With buttons ▲/▼it can be
select one of these options
Pr1
Program 1 (factory setting)
Pr2
Program 2
Pr3
Program 3
Pr4
Program 4
Pr5
Program 5 (from firmware -01)
Pr6
Program 6 (from firmware -01)
°C
Resolution 1°C
0.1C
Resolution 0,1°C
°F
Resolution 1°F
With press
„Set“ the device reboots. For a new selection this must be repeated.
Press for
≥ 2 s
Reset Relay locked switching
Press for ≥ 4 s
Display program number (from firmware -01)
Press for
≥ 10 s
Display firmware version
8, 9, 10 LEDs Sensor
ON
Temperature value of this sensor is being displayed
Flash
Error of displayed sensor, associated alarm LED lights
2 or more LEDs Sensor ON
Warmest of those sensors in display
Alternating illumination
All sensor values are displayed for 2s each consecutively
ON in programming mode
Selected sensor (8, 9, 10) or sensor error acts on this alarm
Hints
:
-
Longer pushing of
▲/▼ accelerates the changes in the display
-
Button
▲/▼ pushed at the same time sets actual programmed parameter to zero