Auto Power Off
When the function switch isn
’t rotated and the button of the instrument isn’t pressed
after 15 minutes, the instrument will turn off the power (in the sleeping). In this mode if you
press any button or rotate the function switch, the instrument will work (be awaking).
When you rouate the function switch to turn on the power, the auto power off
symbols
“
” will show the LCD display. Before instrument will auto power off, it
announce you by sound five beep
Battery replace:
WARNING
: To avoid electric shock, disconnect the test leads from any
source of voltage before removing the battery cover.
1.Turn power off and disconnect the test leads from the meter.
2.Open the rear battery cover by removing one screw using a screwdriver.
3.Insert the battery into battery holder, observing the correct polarity.
4.Put the battery cover back in place. Secure with the screws.
WARNING:
To avoid electric shock, do not operate the meter until the
battery cover is in place and fastened securely.
NOTE:
If your meter does not work properly, check the fuses and batteries to replace.
Replacing the fuses
WARNING
: To avoid electric shock, disconnect the test leads from any source of
voltage before removing the meter cover.
make sure that they are still good and that they are properly inserted.
1. Gently remove the old fuse and install the new fuse into the holder.
2. Always use a fuse of the proper size and value (0.5A/250V fast blow for
the 400mA range .
3. Replace and secure the rear cover, battery and battery cover.
WARNING
: To avoid electric shock, do not operate your meter until the fuse
cover is in place and fastened securely.
.
www.zicotech.com