background image

RESISTANCE MEASUREMENTS 

WARNING:

  To avoid electric shock, disconnect power 

to the unit under test and discharge all capacitors before 

taking  any  resistance  measurements.  Remove  the 

batteries and unplug the line cords. 

1.

 

Set the function switch to the 



 position. 

2.

 

Insert  the  black  test  lead  banana  plug  into  the  negative 

(COM)  jack  and  the  red  test  lead  banana  plug  into  the 

positive 

 jack. 

3.

 

Touch the test probe tips across the circuit or part under 

test. It is best to disconnect one side of the part under test 

so  the  rest  of  the  circuit  will  not  interfere  with  the 

resistance reading. 

4.

 

Read the resistance in the display. The display will indicate 

the proper decimal point, value and symbol.

 

CONTINUITY CHECK

 

WARNING: 

To  avoid  electric  shock,  never  measure 

continuity  on  circuits  or  wires  that  have  voltage  on 

them. 

1.

 

Set the function switch to the                  position. 

2.

 

Insert  the  black  lead  banana  plug  into  the  negative  (-) 

jack  (COM)  and  the  red  test  lead  banana  plug  into  the 

positive (+) jack (

). 

3.

 

Press  the            button  until  the            symbol  appears 

in the display. 

4.

 

Touch the test probe tips to the circuit or wire you wish 

to check.   

5.

 

If  the  resistance  is  less  than  approximately  30

,  the 

audible signal will sound. The display will also show the 

actual resistance. 

DIODE TEST 

WARNING: 

To  avoid  electric  shock,  do  not  test  any 

diode that has voltage on it. 

1.    Set the function switch to                  position. 

2.    Press  the            button  until  the            symbol  appears 

in the display. 

3.   Insert the black test lead banana plug into the negative (-) 

jack  (COM)  and  the  red  test  lead  banana  plug  into  the 

positive (+) jack (

). 

4.

 

 

Touch  the  test  probe  tips  to  the  diode  or 

semiconductor  junction  you  wish  to  test.  Note  the  meter 

reading 

 

5.    Reverse  the  probe  polarity  by  switching  probe  position. 

Note this reading. 

6.    The diode or junction can be evaluated as follows: 

A.  If  one  reading  shows  a  value  and  the  other  reading 

shows OL, the diode is good. 

B. If both readings show OL, the device is open. 

C.  If  both  readings  are  very  small  or  0,  the  device  is 

shorted. 

 

NOTE:

    The value indicated in the display during the diode 

check is the forward voltage. 

 

FREQUENCY MEASUREMENT

 

1.    Set the function switch to the FREQ position. 

2.   Insert the black test lead banana plug into the negative (-) 

jack  (COM)  and  the  red  test  lead  banana  plug  into  the 

positive (+) jack (F). 

3.    Touch the test probe tips to the circuit under test. 

4.    Read  the  frequency  in  the  display.  The  digital  reading 

will indicate the proper decimal point, symbols (Hz, kHz) 

and value. 

 

CAPACITANCE MEASUREMENTS 

WARNING:

  To avoid electric shock, disconnect power 

to the unit under test and discharge all capacitors before 

taking  any  capacitance  measurements.  Remove  the 

batteries and unplug the line cords. 

1.

 

Set the function switch to the CAP position. (“nF” and a 

small value will appear in the display).   

2.

 

Insert the black test lead banana plug into the negative (-) 

jack  (COM)  and  the  red  test  lead  banana  plug  into  the 

positive (+) jack (CAP). 

3.

 

Touch  the  test  leads  to  the  capacitor  to  be  tested.  The 

display will indicate the proper decimal point, value and 

symbol. 

 

 

 

 

 

Summary of Contents for ZI-6940

Page 1: ...When using the probes keep your fingers behind the finger guards on the probes Measuring voltage which exceeds the limits of the multimeter may damage the meter and expose the operator to a shock haza...

Page 2: ...pushbutton SYMBOLS AND ANNUNCIATORS Continuity BAT Low Battery Diode DATA HOLD Data Hold AUTO AutoRanging AC Alternating Current or Voltage DC Direct Current or Voltage SPECIFICATIONS The instrument...

Page 3: ...0k 2 0 of rdg 3 dgts Input Protection 250V dc or 250V ac rms Capacitance Auto ranging Range Resolution Accuracy 40 00nF 10pF 5 0 of rdg 7 dgts 400 0nF 0 1nF 4 000uF 1nF 3 0 of rdg 5 dgts 3 0 of rdg 5...

Page 4: ...the indicator The indicator HOLD will be appear in the display 2 Press the DATA HOLD button to return to normal operation 3 Press the HOLD button for 2 seconds to BACKLIGHT button it will turn on or...

Page 5: ...ion switch to the A position and insert the red test lead banana plug into the 10A jack 5 Press the MODE button until DC appears in the display 6 Remove power from the circuit under test then open up...

Page 6: ...appears in the display 3 Insert the black test lead banana plug into the negative jack COM and the red test lead banana plug into the positive jack 4 Touch the test probe tips to the diode or semicond...

Page 7: ...or AC position using the function select switch 3 Press the trigger to open jaw Fully enclose one conductor to be measured 4 The meter LCD will display the reading REPLACING THE BATTERY WARNING To avo...

Page 8: ...er 4 Always use a fuse of the proper size and value 0 5A 250V fast blow for the 400mA range 20A 250V fast blow for the 10A range 5 Put the cover Insert the screw and tighten it securely WARNING To avo...

Reviews: