16
Guidance and manufacturer's declaration – electromagnetic
immunity –for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of the Infrared
Thermometer should assure that it is used in such an environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV
contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Guidance and manufacturer´s declaration – electromagnetic
immunity –for EQUIPMENT and SYSTEM that are not
LIFE-SUPPORTING
Guidance and manufacturer´s declaration – electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of the Infrared
Thermometer should assure that it is used in such an environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
Portable and mobile RF
communications quipment
should be used no closer to
any part of the
Infrared Thermometer,
including cables, than the
recommended separation