ZEUS-100/80 ZO-OM-008
REV : 08 12
ZERONE
②
Make a short cut of the current path between Active Electrode and Dispersive.
4. Follow-Up Ways for Storage and Maintenance
- When removing a series of codes, don
’
t unplug them at a time.
- Don
’
t screw, bind or stack the cables.
- After performing an operation, sterilize all accessories and then store them.
- Turn off the power after using equipment and remove the power plug and store it
when leaving the office or not using for a long time.
①
About Cleaning
- Clean the main unit and foot switch with soft fabric which is wet with warm water
or alcohol at least once a month. If you use the material is designated cleaning, you
have to ask chemical manufacturer regarding effect of microbe.
Clean its external appearance with non-ignitable, non-explosive materials.
Don
’
t use corrosive, easily-grinding materials that can damage it such as lacquer,
thinner, ethylene and oxidizing agents, etc. Don
’
t let the liquid into equipment.
- You should use 70% isopropyl alcohol or ethylalcohol when you clean accessories.
②
About Sterilization
- Use the accessories that can be sterilized after sterilizing them.
Accessory
Sterilization
Auto Clave
Reusable Bipolar Forceps
- Gravity Displacement (121
℃
(250
℉
))
: 30minutes
- Drying time
: 20 minutes
Reusable Electrodes Tips
Reusable Monopolar Handle
- Gravity Displacement (121
℃
(250
℉
))
: 20minutes
- Drying time
: 20 minutes
①
During sterilization, don
’
t increase temperature and pressure more than the standard,
which can cause damage to the accessories.
②
Don
’
t sterilize disposable accessories.
Check the followings before using this product.
Warning