background image

Q

Q

Q

QUICK SETUP GUIDE

UICK SETUP GUIDE

UICK SETUP GUIDE

UICK SETUP GUIDE

 

 

 Zensol Automation Inc.                                    

www.zensol.com

 

 

27 

How to use this plan: 

 
The test plan has only 4 activated tests. They all verify one input at the time. 
 
To execute the tests, click the 

Execute Test

 button. To change tests, use the 

blue arrows

 

 

 
Test 1– AC Clamp Test. 

Plug  a  small  wire between  the two  “Current”  connectors.  The  AC  clamp must pass  through the wire  and be 
plugged in the input #1, as seen in the image below. 
 
You are now ready to execute the test. 

 
 
 
 
 
 
 
 
 
 
 

 
Test 2, 3 and 4 – Accelerometer Tests. 

 
Screw  an  accelerometer  to  the  chassis  of  the  TAP-4  to  the  designated  area.  Then,  plug  the  accelerometer 
cable into the corresponding input you want to test. 
You are now ready to execute the test. 
 

 
 
 
 
 
 

Summary of Contents for TAP-4

Page 1: ...Zensol Automation Inc www zensol com COMPUTERIZED TEST EQUIPMENT ZENSOL ON LOAD TAP CHANGER ANALYZER OpenZen Universal Data Recording Software Quick Setup Guide TAP 4 August 13th 2013...

Page 2: ...INDOWS VISTA 7 8 9 STEP 3E CONTENTS OF THE OPENZEN FOLDER 10 STEP 4 COMMUNICATION USB DRIVER INSTALLATION OR FIBER OPTIC INSTALLATION 11 STEP 4A FIRST INSTALLATION 11 STEP 4B IF THE FOUND NEW HARDWARE...

Page 3: ...TUP GUIDE QUICK SETUP GUIDE Zensol Automation Inc www zensol com 2 STEP 1 REQUIRED HARDWARE TAP 4 Power cable USB cable Ground cable Accelerometer with calibration card 10 32 cable BNC cable male fema...

Page 4: ...p 1a Accessories Current Clamp In input 1 plug the AC DC current clamp and install it on a current wire from the tap changers motor as shown on the following picture Accelerometers In the other inputs...

Page 5: ...RAW Signals 1 Current 2 Accelerometer 1 3 Accelerometer 2 4 Accelerometer 3 Envelopes Signals 100 Current 101 High Frequency Envelope Accelerometer 1 102 Low Frequency Envelope Accelerometer 1 103 Hig...

Page 6: ...QUICK SETUP GUIDE QUICK SETUP GUIDE QUICK SETUP GUIDE QUICK SETUP GUIDE Zensol Automation Inc www zensol com 5 Step 2b Connections to the tap changer...

Page 7: ...TURE Step 3a Downloading the application from the Zensol website Go to www zensol com and in the products menu select TAP 4 OLTC Analyzer In the product page of the TAP 4 scroll down to the CUSTOMERS...

Page 8: ...n the icon for the file you downloaded Select Extract to note this example uses WinZip other extractors may vary in their method or file for latest version and date Select the folder of your choice an...

Page 9: ...not be executed Double click on the icon for the file you just downloaded or file for latest version and date Click the Extract button note this example uses WinZip other extractors may vary in their...

Page 10: ...nd installing on Windows Vista 7 8 Right click on the icon for the file you just downloaded exact file name will depend on version downloaded Enter the destination folder where you want the files deco...

Page 11: ...tains the program the USB driver a reference file cbawxwin cba default test plans default wcf calculation files cbacal cal Calibration folder This contains the calibration file Docs folder This folder...

Page 12: ...lug the TAP 4 into a free USB port In this window select Install from a list or specific location and click Next Now select Search for the best driver in these locations and Include this location in t...

Page 13: ...is not launched Click Windows button bottom left of your desk then control panel in the right row This window appears Select Device Manager then click right on USB Serial Port Choose to Update Driver...

Page 14: ...feature of the driver the virtual COM port which is not necessary for our application Go into the device manager as seen in previous page and locate your device under the USB branch Double click on i...

Page 15: ...need to re install it from the Device Manager of Windows located in the systems properties To get to the device manager Go into the control panel of Windows Double click on the icon System Go into the...

Page 16: ...p 5a Language configuration Run the executable file OpenZEN TAP EXE or file of latest version The first time it is run you will be prompted to choose for the default language Select English and click...

Page 17: ...IDE UICK SETUP GUIDE Zensol Automation Inc www zensol com 16 Step 5b Main window Toolbar Number and Name of the Test Number and Name of the Page File name Signals information Calculus related to data...

Page 18: ...ode Test configuration Check setup Previous graphic page Open plan Informations Previous Test Next graphic page Save raw data only Analog inputs Next Test Examine time amplitude Save Hydro Quebec enve...

Page 19: ...SETUP GUIDE UICK SETUP GUIDE Zensol Automation Inc www zensol com 18 Step 5c Data structure RAW DATA DEFAULT_CHECK_TAP4 wdt EXCEL TAP Report_English xls TEST PLAN DEFAULT_CHECK_TAP4 wcf CALCULUS CBAC...

Page 20: ...ion Inc www zensol com 19 Step 5d Calibration checkup Once the Verification button is pressed the software connects to your instrument reads the calibration data and applies it to your test plan It ha...

Page 21: ...on A welcome window will appear asking you what you want to do Select the test plan generator and click continue In this screen you can write down all the necessary information about your tests If you...

Page 22: ...s Once you have verified the calibration in the previous step you can scale this calibration to the real values you measure A good habit to have is to save the test plan once our calibration data is e...

Page 23: ...nputs The same thing is done with the accelerometers inputs The unit of measurement is in G 0 Note If you don t know precisely the calibration value we recommend writing 100 mv g for a 50g max scale F...

Page 24: ...st launched would be 8 If you have configured your tests with the TRIG once the test is launched you can change taps on your tap changer when you see the Measure LED light up on the front panel of the...

Page 25: ...lete click on the Save button in the main window s button bar It is also possible via the menu under File Save Data as To save only the envelopes smaller files you only have to choose the option Save...

Page 26: ...r examples of OpenZen TAP 2 In the window Processing load the excel report that you want as shown below 3 Click on the last red button to do the export 4 Excel file specified in the processing window...

Page 27: ...test plan if you want to verify your hardware before real tests In the main window s button bar click the Assistant Mode button The welcome window will appear asking what you want to do Select Load an...

Page 28: ...change tests use the blue arrows Test 1 AC Clamp Test Plug a small wire between the two Current connectors The AC clamp must pass through the wire and be plugged in the input 1 as seen in the image be...

Page 29: ...Q Q Q QUICK SETUP GUIDE UICK SETUP GUIDE UICK SETUP GUIDE UICK SETUP GUIDE Zensol Automation Inc www zensol com 28 STEP 7 SYMBOLIC CIRCUIT DIAGRAM...

Reviews: