Part List
Fig. 4
Position
Description
Quantity
1
Base plate
1
2
Spacer plate
1
3
Hexagon bolt
2
4
Washer
2
5
Stub
2
6
Nut (Lower)
2
7
Nut (Upper)
2
8
Washer
2
9
Hexagon socket head counter sunk bolt
2
10
Top plate
1
11
Lower bearing
1
12
Rocker pin support
13
Rocker pin
1
14
Earth cable
1
15
Hexagon bolt + washer
2
Use
Routine maintenance includes the optical inspection of bolts, plates and force bearing parts, as well as
the removal of dirt, dust and debris built up near or between the load cell and/or mount in order to let it
move freely.
Finally
A load cell will perform within specifications until the safe load limit or safe side load limit is passed.
Beyond this point, even for a very short period of time, the load cell will be permanently damaged. The
load cell may physically break at the ultimate load limit.
When not working properly remove load cell with care and attach a label with comments relating to the
problem or mode of failure. Return also a copy of the Certificate of Calibration with the load cell if
available.
Don't cut cable at the gland to facilitate removal - please - we cannot test load cells without cables!