background image

 

III.

 

Visual examination of the sample 
 

1.

 

Press “Locate” in the main microscope menu and “Online” 

2.

 

You can change objectives automatically on the small touch screen attached on the right side 
of the microscope by selecting “microscope” menu and “objectives” 

3.

 

Please make sure the chosen objective corresponds to the one shown in the software. If this is 
not the case please restart the system. (Otherwise your pinhole and pixel sizes as well as the 
sectioning thickness are all incorrect.) 

4.

 

If you wish to view your sample in brightfield, press the TL button in the “Locate” menu 

5.

 

For visual examination of fluorescence, press the buttons DAPI, GFP, RFP or FarRed 

6.

 

If sample appears too dim, increase light intensity by pressing the ‘Colibri 5/7’ button and 
adjust the LED intensity slider. 

7.

 

Before switching to acquisition mode, close transmitted and reflected light shutters by 
pressing ‘off’ and turn off the LED light source by clicking on ‘Colibri 5/7’ and pressing 
‘Deactivate all LEDs’. 
 
 
 

IV.

 

Acquiring confocal images 

 

 

Select the ‘Acquisition’ tab in the main menu of Zen blue  

 

Go to “Smart Setup” choose your dyes (set the LUT) and configuration, you can choose 
between simultaneous (fastest), Best compromise line scanning mode (Smart) and sequential 
(best signal) scan, press ”Apply”.  

 

If you use ‘Smart setup’, please make sure that you DO NOT have any track using “Ch2”, 
which is the Airyscan GaASP detector!
 You’ll have to manually switch the channels that are 
assigned to it to a different detector: low wavelengths to ‘Ch1’ and high wavelengths (from 
mCherry on) to ‘Ch3’. Due to the arrangement of dichroic mirrors in the scanhead, it won’t 
work the other way round! 

 

If you are using the detector for transmitted light (ESID): activate (ESID) represented in the 
light path scheme. Adjust intensity in the ESID Gain, then adjust digital offset and digital gain 
of the ESID detector in order to “stretch” the histogram and use the whole dynamic range of 
the detector.  

Reviews: