LSM 880
Center Screen Area / Image Containers - Display and Image Analysis
ZEISS
10/2014 V_01
000000-2071-464
517
If you press the
Settings
button, the settings
menu will pop up (Fig. 723).
In
Map Selection
you can determine the map
type with the
Map type
pull down menu (Fig.
724). You can choose between the
Number
particles
and the
Diffusion coeffic. (
µ
m x
µ
m
/s)
parameter maps, that will display the relative
molecule numbers –please note that if average
subtraction is performed the information on
absolute numbers is lost - and diffusion
coefficients of the mapped areas.
You can also display the informative pixels by
choosing
Inf.Pix
. This will display a map that
indicates, how many pixels contributed
meaningfully to the correlation before dropping
to zero.
If the
Auto Scale
check box is activated, the
maps are scaled automatically between the
highest and lowest value. If unchecked you can
enter in the
Min
and
Max input boxes
own
minimum and maximum values. All values lying
between will be accordingly scaled. Note that all
values lying outside the minimum and
maximum will be displayed as black in the
maps. This option is helpful in eliminating
extreme outliers.
Fig. 723
The Map settings menu
Fig. 724
Map type pull down menu