Exchanging the Sample Mounting Frames
The stage proper with the incubation unit can receive sample mounting frames with different formats.
Dismounting:
1.
Remove the incubator lid.
2.
Press the frame against the incubator spring
using the two handling screes and lift the frame
up from the other side.
3.
Remove the frame.
4.
Insert the incubator lid.
Mounting:
1.
Remove the incubator lid, see section 1.13.3.
2.
Take a suitable frame (Fig. 17/
1
) using the two
handling screws and press with one side against
the spring of the incubator stage (Fig. 17/
2
).
3.
Make sure the mounting frame is tightly inserted.
4.
Insert the incubator lid.
1
Mounting frame with two handling screws
2
Spring
3
Incubation stage receptacle
Fig. 17
Exchanging of Sample Mounting
Frames