Database Management
11-11
patient database from a USB storage device, floppy disks, or from your office network if you have
licensed the networking feature of your HFA II-
i
.
summarizes the pathway between
database failure and problem resolution. The following pages lead you through the necessary steps,
for each of the options discussed.
When faced with a hard drive database problem, your options are:
A. Attempt to fix immediately (FIX NOW).
B. Delay attempt to fix problem, continue patient testing, and save results to a USB storage
device only (USB ONLY).
C. When time permits, engage the rebuild utility (REBUILD HARD DISK DATABASE).
D. Restore database from a USB storage device, from your network file server, or backup
floppy disks if above options fail (BACK UP/RESTORE).
CAUTION: If a hard drive database failure occurs while attempting to save test results,
do not recycle power at this time. This action will cause you to lose the results from
the test just completed. Always elect either to fix the problem immediately, or to save
results to a USB storage device temporarily until you have time to address the
database problem.
Figure 11.2 Pathways for Resolving a Hard Disk Database Failure – An Overview
provides an overview of different methods for dealing with hard disk database failures.
More detailed information, including step-by-step instructions to perform these steps, follows.
Hard Drive
Database
Failure Detected
Attempt to fix?
Stop Testing
Choose FIX NOW
Continue Testing
Choose save to USB ONLY
Delay fix?
Recycle power
Attempt to fix?
Try REBUILD
Unsuccessful?
HARD DISK DATABASE
Successful?
Copy files
from USB storage device
to hard disk
Unsuccessful?
Unsuccessful?
Delay fix?
Attempt new fix?
Attempt new fix?
Try REBUILD
HARD DISK
DATABASE
Successful?
Resume Testing
Unsuccessful?
RESTORE patient
database from
USB storage device
floppy disks,
Successful?
Copy files
from USB storage device
to hard disk
Successful?
Save to USB
ONLY
Resume Testing
or your network file
server
RESTORE patient
database from
USB storage device
floppy disks,
or your network file
server
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......