Test Reliability
6-5
In addition to a high false positive finding, trigger happy patients often show threshold results that
are abnormally high. An example of this phenomenon is shown in
. Any individual point
result of 40 dB or greater indicates a hypersensitive result which can only be due to patient
overreaction or guessing when pressing the patient response button. It is best to note the results
early in the exam and start the test over rather than to allow the test to run to completion and be
completely invalid.
Gaze Graph
Fixation Losses
False Positive
False Negative
Abnormally high results
Figure 6.1 Sample Printout Showing Poor Reliability
indicates a very unreliable patient. A high number of fixation losses, false
positive errors, and false negative errors have been recorded. Poor fixation is indicated by the Gaze
Graph. Also note the number of points where threshold results are 40 dB or higher.
Summary of Contents for Humphrey II-i series
Page 1: ...User Manual Humphrey Field Analyzer II i series System Software Version 5 1...
Page 8: ...Humphrey Field Analyzer II i series User Manual 2660021145640 A vi...
Page 74: ...General Operation Humphrey Field Analyzer II i series User Manual 2660021145640 A 2 34...
Page 130: ...Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 5 20...
Page 158: ...STATPAC Analysis Printing Humphrey Field Analyzer II i series User Manual 2660021145640 A 7 22...
Page 338: ...Kinetic Testing Humphrey Field Analyzer II i series User Manual 2660021145640 A 13 62...
Page 412: ...Care and Cleaning Humphrey Field Analyzer II i series User Manual 2660021145640 A 15 14...
Page 416: ...Product Specifications Humphrey Field Analyzer II i series User Manual 2660021145640 A A 4...
Page 420: ...Product Features Humphrey Field Analyzer II i series User Manual 2660021145640 A B 4...
Page 428: ...Icon Glossary Humphrey Field Analyzer II i series User Manual 2660021145640 A D 2...
Page 579: ......