Operation
16
626109-9160-000 FixAssist XTR Operating manual V1.0
Below the adjustment of the second axis is described using a probe with
a cube and screw as an example.
6.
7.
Replace the probe on the
cube with a stainless steel
extension.
Release the screw (
1
) of the
cube slightly. The cube can
now be rotated.
1
8.
9.
Release the fixing screws (
2
)
of the angular adjustment
and set the desired angle of
the second axis.
Tighten fixing screws again.
2
10.
11.
12.
13.
Push the prism slide
horizontally, carefully until
the extension is against the
prism without backlash.
Re-tighten the screw of the
cube.
Check the angle once more.
Replace the stainless steel
extension with the probe.
The probe system is set up
and ready for operation.
Fig. 15:
Fig. 16:
Fig. 17: