FixAssist

®

 XTR

Probe presetting device

for probe systems with 
XTR plate

Operating manual

Summary of Contents for FixAssist XTR

Page 1: ...FixAssist XTR Probe presetting device for probe systems with XTR plate Operating manual...

Page 2: ......

Page 3: ...ld in the sense of trademark and brand protection legislation be regarded as unused and therefore usable by anyone Layout illustration photos and text Carl Zeiss 3D Automation GmbH Printed in Germany...

Page 4: ...data 7 3 1 Intended use 7 3 2 Misuse 7 3 3 Scope of delivery 8 3 4 Design 9 3 5 Technical data 10 4 Operation 11 4 1 Inserting probe plates 12 4 2 Configuring the angle around the Z axis 13 4 3 Config...

Page 5: ...e of the causes given below Improper use of the product Incorrect installation start up operation and maintenance of the pro duct or use of spare parts not approved by the manufacturer Failure to obse...

Page 6: ...be able to operate the loading station when in a sitting or upright position Follow the instructions of the work safety legislation 2 2 Structural modifications No modifications additions or conversi...

Page 7: ...e around the Z axis of the coordinate measuring device and around a horizontal axis can be accurately and reproducibly calibrated All of the specifications in these operating instructions must be heed...

Page 8: ...Scope of delivery Product Image FixAssist XTR Tool kit consisting of 1x 1 2 Nm torque screwdriver 2x DG11 hook wrench 2x DG20 hook wrench 1x M5 V2A probe extension 1x angle adjustment device for DG11...

Page 9: ...9 9 Fig 3 FixAssist structure 1 Plate support 2 Centring aid 3 Prism 4 Vertical guide 5 Fixing screw yellow for vertical guide 6 Fixing screw black for angle adjustment 7 Fixing screw for horizontal...

Page 10: ...l data 10 626109 9160 000 FixAssist XTR Operating manual V1 0 3 5 Technical data Condition Value Weight 19kg Dimensions 700mm x 230mm x 515mm LxWxH Ambient temperature 5 40 C Humidity 30 95 non conden...

Page 11: ...red probe system Faults during assembly have a direct impact on your measurement results Always use a stainless steel extension when bolting down extensions or probes to hold them in place Fig 4 Exten...

Page 12: ...holder In this process the alignment plate 2 fits into the groove 3 The two snaps of the probe plate move into the recesses designed to accommodate them 2 3 3 Fix down the probe plate Push the red fi...

Page 13: ...r scale and adjust the desired angle by rotating the angular scale In this process think about in which direction you want your probe system to turn Re apply the wing nuts The angle of the probe plate...

Page 14: ...he probe sphere must be right in the centre of the prism Fig 11 10 11 12 Re apply the screws 5 slightly Check whether the probe sphere is still centred in the prism If the configuration is cor rect ti...

Page 15: ...nfigured 1 Insert the centring aid 1 into the prism slide 1 2 3 4 5 Release the fixing screw of the vertical and horizontal guide Push the centring aid carefully both vertically and horizontally until...

Page 16: ...lightly The cube can now be rotated 1 8 9 Release the fixing screws 2 of the angular adjustment and set the desired angle of the second axis Tighten fixing screws again 2 10 11 12 13 Push the prism sl...

Page 17: ...rements for all materials Dusty parts are damp cleaned never wipe in dry condition It is necessary to use a mild detergent and soft cleaning materials cloth sponge or chamois leather The following are...

Page 18: ...nal waste disposal regulations 6 2 Notes on storage If the FixAssist XTR is put into storage it must be stored in a dry and dust free location The recommended storage temperature is 5 C to 40 C 6 3 Tr...