5 First Operating Steps | 5.4 Acquiring an Image
ZEISS
4. Turn the knob of the mid-column aperture
changer to select the 20 µm aperture.
5. In the
Apertures
tab of the SEM Controls panel, select the
20.00 µm
from the
Aperture
Size
drop-down list.
à
The 20 µm aperture is set.
5.4.9.2 Setting EHT and Probe Current
Procedure
1. In the SEM Controls panel, select the
Gun
tab.
2. Double-click the
I Probe
readout and enter
100
.
à
The probe current is set to 100 pA.
3. Double-click the
EHT Target
readout and enter
20
.
à
5.4.9.3 Selecting the SE Detector
Procedure
1. In the SEM Controls panel, select the Detectors tab.
2. In the Detectors section, select
Signal A = SE2
from the
Signal A
drop-down list.
INFO:
We recommend using the SE detector to
obtain the first image. This detector provides a
good signal-to-noise ratio even at long working
distances.
74
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006