5 Description | 5.3 Technical Data
5.3 Technical Data
Specification
Capacity
0.6 l
Size of dewar assembly
overall height 334 mm, maximum diameter
152 mm
Cooling agent
liquid nitrogen (LN
2
)
Cooling-down time
30 minutes to reach -130°C
45 minutes to reach -170°C
Temperature of cooling plate
down to -180°C
Standing time
After refilling the dewar the temperature of
the cooling plate stays below -130°C for
10…12 hours.
Warming-up time
At regular system vacuum:
With a refilled dewar it takes about 2 days to
reach room temperature again (36 hours to
reach 20°C).
CryoWarmUp macro:
A macro allows to reduce the warming-up
time. With a refilled dewar it only takes
about 6 hours to reach room temperature
again (4.5 hours to reach 20°C).
System vacuum
Improvement of an approximate factor of 2
5.4 System Requirements
§
Airlock
§
Liquid nitrogen, has to be provided by the customer